Remote Alarm System; Cycle Complete; Power Failure - Thermo Scientific MAXQ SHKE435HP Operating And Maintenance Manual

High performance console incubated orbital shaker
Table of Contents

Advertisement

Section 2
Operation

Remote Alarm System

Cycle Complete

Power Failure

2-14
Orbital Shaker
Most of the alarm states described previously (see Table 2-2) can alert a
remote alarm monitoring system through an internal SPDT relay
connected to an RJ-11 jack on the rear of the shaker cabinet. Refer also to
'Connect the Remote Alarm' in Section 1. For the convenience of the
laboratory, these remote alarms can be individually turned on or off. Any
of the remote alarms set to On will activate the internal relay.
Note The remote Overtemp Shutdown, Platform Stalled and Check Fuse
alarms cannot be deactivated.
To set the remote alarms to
On or Off, open the
Configuration menu by
pressing the down arrow, up
arrow, and the Silence button,
and then the down arrow three
times, in the sequence shown at
right. The screen shown at right
will appear on the display.
Press the Speed button beneath Remote (Rmte). The alarms will be shown
in the following sequence.
Toggle the Cycle Complete alarm with either the up (On) arrow or the
down (Off) arrow. Pressing Temp
button beneath Next advances the
display to the next alarm, saving the
shown Cycle Complete setting to
memory.
If no buttons are pressed, the display automatically returns to the
Operating Screen after about 15 seconds, saving the selection to memory.
Toggle the Power Failure alarm with
either the up (On) arrow or the down
(Off) arrow. Pressing the Temp button
beneath Next advances the display to the
next alarm, saving the Power Failure setting to memory.
If no buttons are pressed, the display automatically returns to the
Operating Screen after about fifteen seconds, saving the selection to
memory.
s
Cycle Complete:
Actual
Setpoints
Actual
Power Failure:
Setpoints
On
Next
On
Next
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents