Install Test Tube Holders; Rs-232 Interface Connector - Thermo Scientific MAXQ SHKE435HP Operating And Maintenance Manual

High performance console incubated orbital shaker
Table of Contents

Advertisement

Install Test Tube
Holders (continued)
RS-232 Interface
Connector
Thermo Scientific
To install the Test Tube Rack Holder onto
the shaker platform, remove the rack and
rotate the swing-bed of the holder 90° by
pulling the knobs of the locking pins on
either end of the holder outward. The pins
are locked outward by turning the knob
1/4-turn (Figure 1-14). Attach the tray to
the platform with the screws provided.
The MaxQ Console Orbital Shaker is equipped with an RS-232 Serial
Communication Interface for the remote transmission of data. An RJ-11
telephone style connector is located on the left side of the incubator. A
cable with RJ-11 plugs and an RJ-11 to DB-25 adapter are required. Refer
to Figure 1-18 for connector locations on the shaker back panel. Figure 1-
15 identifies the RS-232 and Remote Alarm pin contacts.
The data is "dumb terminal" formatted, which permits interfacing with
either a computer or a serial printer.
ALARM CONTACTS
30V MAX / 1 A MAX
(2) COM
(3) N.O.
Figure 1-15. Remote Alarm and RS-232
Connectors
Figure 1-16. Remote Alarm Output to Screw Terminal Connection
RS-232
INTERFACE
(2) RXD
(3) SIG. GRD
(4) N.C.
Figure 1-14. Rotate Rack
(4) TXD
Orbital Shaker
Section 1
Installation
1-9