Checking Input Range

Checking Input Range
Input range
Input range
description
3-4
Shop for Power Metering products online at:
The Input Range screen is displayed by pressing the CkRng key on the Scope mode
screen. Input Range shows the detected range for all channels. The number after the
status shows the percentage of the input of the instrument's full range.
The input range can only be viewed, not changed. Press OK when done viewing the
input range.
The following table describes the messages that may register on the Input Range
screen. The instrument is able to detect the input range for voltage and current channels
A, B, C, and D.
Input Range
OK
OVER
UNDER
N/A
CLIP
If input range is OVER, UNDER, or CLIP, then the rms readings will not be accurate.
Using CTs below 10% of Full Scale may also cause inaccuracies in amplitude, phase
and harmonic distortion readings.
PowerMeterStore
www.
MARK205
Description
Within Range
Over Range
Under Range
Channel Disabled
Clipping
1.800.561.8187
.ca