Maintenance

Cleaning

Required for cleaning is a cleaning adapter (item no: 50133431).
Use the following disinfecting agents for cleaning:
Cleaning Solution, 1 syringe, item no.: CMX 25.
Always wear protective gloves when handling the cleaning solution.
Always wear safety goggles when handling the cleaning solution.
Smart2Pure
68
• The Smart2Pure systems can only be disinfected as a 'stand-alone' process.
• If the system was longer time not in operation a cleaning must be performed.
• Cleaning should be performed at regular intervals, such as when the ultrapure
cartridge is exhausted and must be changed, or when there is bacteria present the
pure water.
• Depending on the version of your Smart2Pure system (with internal 5L tank
Smart2Pure 3/6 or Smart2Pure 12 with external tank), a complete cleaning will
take between 1 to 5 hours.
• To ensure effective cleaning of your system the internal 5L or external tank must
be filled completely with water.
• There is no need to worry should you notice a chlorine smell during the entire
cleaning process, as there is no risk of the limit for chlorine gas in closed rooms
being exceeded.
If your skin should come into contact with a chlorine product, rinse it immediately
with ample, fresh water. If your eyes come into contact with the disinfecting agent,
rinse them immediately with ample, fresh water and contact a physician at once.
Observe the information provided on the data sheet supplied with the CLEANING
Solution to avoid any possible health hazards.