Connecting An External Tank To The Smart2Pure 12 System - Thermo Scientific Smart2Pure 3 Operating Instructions Manual

Barnstead smart2pure series ultrapure water system
Table of Contents

Advertisement

7
Installation
Connections of the Smart2Pure 12 system

Connecting an external tank to the Smart2Pure 12 system

Step
Action
1
Place the external 30L or 60L pure water tank in the
direct vicinity of the Smart2Pure 12 system.
2
Use the AD 1/4" tube to connect the tank back flow
connector on the pure water tank with the tank back flow
connector on the Smart2Pure 12 panel.
3
Plug in the AD 1/4" tube to connect the tank flow
connector on the pure water tank to the tank flow
connector on the Smart2Pure 12 panel.
Smart2Pure
34
• The external pure water tanks are supplied only with the Smart2Pure 12 system.
• The external pure water tanks are available in two different sizes:
• 06.5040 pure water tank 30L
• 06.5070 pure water tank 60L
• All connectors for connecting the 30L or 60 L pure water tank to the
Smart2Pure 12 system have the same position.
• If you ordered the sterile overflow for pure water tanks, connect the sterile
overflow as described in the separate installation guide. If not use a AD 8mm/
0.31 inch tube to connect the tank overflow to drain as described in this
chapter.
• A tank vent filter has to be ordered separately (see
"Accessories"
Figure
Tank back flow
connection tank
Tank flow
connection tank
on
page
87).
Tank back flow
connection
Smart2pure 12
Tank flow
connection
Smart2pure 12