Step
Action
4
Locate the cable for level control on the pure water tank
and take it into the level control connector of the
Smart2Pure 12 panel.
5
Use a AD 8mm/ 0.31 inch tube to connect the tank
overflow fitting connector with the drain. Route the AD
8mm/ 0.31 inch tube with a free gravity fall to the drain.
(see
"Illustration of drain"
6
Open the feedwater tap and check that there is no leakage
on the connectors and tubes.
Illustration of drain
Illustration of drain siphon for inserting the rinsing water
and concentrate outlet tube.
Shorten the rinsing water and concentrate outlet tube to
the required length and route it to an atmospherically
vented drain.
The tubes that run from the Smart2Pure system and the
external tank to the drain must be routed with a
downward slope and without any kinks or restrictions, as
this would result in backing up of the draining water.
If a standard drain siphon is in place, the ends of the tubes
must be located at least 20 mm/ 0.75 inch above the
drain. Attach the tubes in such a manner that they remain
in their position.
Thermo Scientific
on
page
35)
Connections of the Smart2Pure 12 system
Figure
Level control
connection
system panel
Level control
cable
Overflow tube
Rinse water tubing
Minimum distance
20 mm/0.75 inch
7
Installation
Tank
Tank overflow
fitting
AD 8mm/ 0.31
inch tube
Drain
Smart2Pure
35