12
System control
Menu
Mono/dual measurement mode
Step
Action
1
Press the „"Menu" button once then the
display shows you the last adjustment
measurement mode (green LED lights up).
The first line shows the selected unit of
measurement resistance value (factory
adjustments in M ·cm).
2
Use the "up/Quit" and "down" button to
choose the resistance value for measuring
mode (μS/cm or M ·cm) and than choose
the mono measurement mode "c" or dual
measurement mode "ct".
Smart2Pure
58
The menu consists of three sub-points:
1. Switching to dual mode together with selection of the measurement unit for the
conductivity measurement.
2. Setting the product water conductivity limit value.
3. Setting the temperature limiting value.
This following combinations are possible in the Mono/dual measurement mode:
• adjusted resistance value M ·cm: "c" and "ct"
• adjusted resistance value μS/cm: "c" and "ct"
Mono measurement mode:
• In this mode only the conductivity is displayed. The temperature can be read when
the "Enter" button is pressed.
Dual measurement mode:
• In this mode, temperature and conductivity are shown alternating in a 2-second tact.
The LED for the unit switches with the display.
Picture
Display shows
Display
First line
or
ct
18.2
μS/cm
°C
M ·cm
Display shows
or
c
Thermo Scientific