This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents. Thermo Scientific GenPure...
Please consult your local sales representative for details. Warranty Thermo Electron LED GmbH warrants the operational safety and functions of the Thermo Scientific Barnstead Ultrapure Water Systems only under the condition that: • the system is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions, •...
Risk of electric shock! Electrical work on the system is only to be carried out by qualified personnel. Protective conductor connection. Connect the power supply to an electrical socket with a protective connection. Indicates a situation where protected gloves or clothing is needed. Indicates a situation in which protective goggles must be worn. Thermo Scientific GenPure...
Page 6
Additionally, the ultrapure water system is in compliance with many other international standards, regulations and directives not listed here. Should you have any questions regarding compliance with national standards, regulations and directives applicable for your country, please contact your Thermo Fisher Scientific sales organization. GenPure Thermo Scientific...
Contents • “Examination on receipt” page 6 • “Complaints” page 6 • “Packing for return shipment” page 6 Do not pull the plastic foil over your head. Risk of suffocation. Use the plastic foil only for packaging. Thermo Scientific GenPure...
• Prior to send back a operated system, empty the water and dry the system and take out the cartridges. • Pack the filter cartridges into a bubble wrap and/ or packaging foam include it with the package of the ultrapure water system. GenPure Thermo Scientific...
Safety precautions Observe these safety precautions for your own safety. Thermo Scientific Barnstead Ultrapure Water Systems are modern water purification systems intended solely for the treatment of potable water or water of ASTM Type II quality. The water it produces is not fit for drinking.
Page 14
UV-lamp glass. • Wear safety goggles when working with cleaning solution. • If your eyes come into contact with a chlorine product, rinse them immediately with ample, fresh water and immediately contact a physician. GenPure Thermo Scientific...
Page 15
• Do not use oxidative cleaning agents when cleaning the system. They would cause damage to it. • Proceed as follows when the system has a defect: • Switch the system off and unplug the system from outlet. • Shut off the feedwater supply. • Contact your local service organization. Thermo Scientific GenPure...
30min is recommended. The procedure for the filter can be repeated up to 5 times. Transformer-table power pack Catalog no.:50149597 Universal Holder and Universal adapter Catalog no.: 21.0007 Catalog no.: 21.0006 Feedwater connecting kit Catalog no.: 25.0075 GenPure Thermo Scientific...
Page 19
Sub-D extension cable, 25 pin, 5 m or 5.47 yards Catalog no.: 16.0375 PE hose, Ø8mm x 20 m or 0.31 inch x 21.87 yards Catalog no.: 18.0036 Sub-D-extension cable, 9pin, 3m or 3.28 yards Catalog no.: 16.0397 Thermo Scientific GenPure...
The system allows the purification of water into the water categories mentioned in the standards of ASTM 11.01 and ASTM 11.02. The Thermo Scientific Barnstead Ultrapure Water System is designed to be installed and use in the following application areas: - Laboratories for cell biological and biotechnological work with the safety levels L1, L2 and L3.
< 0.2 < 0.2 Performance l/min** up to 2 up to 2 up to 2 up to 2 up to 2 up to 2 * Depends on the feedwater and disinfection ** Depends on the feedwater pressure Thermo Scientific GenPure...
Page 26
52.91 lbs (dry weight) GenPure UV-TOC 24 kg 52.91 lbs (dry weight) GenPure UV-TOC/UF 25 kg 55.12 lbs (dry weight) When the system is operating the system is by the amount of water about 3kg / 6.61 lbs heavier. GenPure Thermo Scientific...
Page 27
26.46 lbs (dry weight) Dimensions and weight xCAD Server, xCAD Client (wall version) Height approx. 655 mm 25.79 inch Width 260 mm 10.24 inch Depth approx. 530 mm 20.87 inch Weight 5 kg 11.02 lbs (dry weight) Thermo Scientific GenPure...
Page 28
DC 24 V, 5.0 A max System connection DC 24 V, 80 W Serial interface RS 232 Protection Class Class II (external SMPS certified as Class I) Electrical connections xCAD/Server 1x SUB-D socket 25 pin 2x SUB-D socket 9 pin GenPure Thermo Scientific...
• Easy access for operation and control of the system. • Water pre treated such as DI, RO or distillation water connection with 3/4 NPT male thread and customer supplied shutoff valve. Thermo Scientific GenPure...
Page 32
The installation area GenPure Thermo Scientific...
O-ring on it. Screw the fitting back to the Fitting system. Feedwater connector unit Only feedwater that has been pretreated by reverse osmosis, ion exchange or distillation is to White O-ring Hose 8 mm o.d be used. Thermo Scientific GenPure...
Page 38
The other end of the 0.31” (8 mm) o.d hose you should connect to the xCAD backflow connector of the xCAD Server. Hose 0.31“ White O-ring Fitting (8 mm) o.d Connector xCAD Server Hose 0.31” (8 mm) o.d GenPure Thermo Scientific...
Page 39
Plug the other end of the 25 pin cable for system control into the xCAD Server 25 pin cable connector. System control If applicable use the RS232 connector (13) to connect the optional data printer. RS232 25 pin connector 25 pin cable system control Thermo Scientific GenPure...
Page 40
Open the feedwater supply and switch the system Rinse about 3 liters through the final filter before use. The system is now ready for use. Feedwater supply Only feedwater that has been pretreated by reverse osmosis, ion exchange or distillation is to be used. GenPure Thermo Scientific...
To wall mount the xCAD Server wall version unscrew the 4 screws (see red arrows in the xCAD picture) of the underside from the xCAD Server Server and remove the wall mount bracket. Screws Wall bracket Thermo Scientific GenPure...
Page 42
Connect the wall mount bracket to the wall by screwing in the 4 supplied screws with a cross Screw screw driver into the wall where you installed the plugs before. Cross screw- driver Wall mount bracket GenPure Thermo Scientific...
Page 43
Server 25 pin cable connector. System control If applicable use the RS232 connector (13) to connect the optional data printer. RS232 25 pin connec- tor xCAD Server 25 pin cable system control Thermo Scientific GenPure...
Page 44
Push each of the quick connectors onto the Outlet Quick cartridge. You will know they are attached when connectors an audible “click” is heard. Inlet Fit the cartridge cover on again. Filter cartridge GenPure Thermo Scientific...
Page 45
Hose 0.31” 8 mm o.d Screw the final filter in counter clockwise direction (see red arrow in the picture) into the 1/4" female thread of the xCAD dispensing valve. 1/4" female thread connection Final filter 0.2 μm Thermo Scientific GenPure...
Page 46
“Mounting the power pack (voltage supply)” page Open the feedwater supply and switch the system The system is now ready for use. Feedwater Only feedwater that has been pretreated by supply reverse osmosis, ion exchange or distillation is to be used. GenPure Thermo Scientific...
Server When the xCAD Client is finish connected to the xCAD Server the xCAD Server must be in operating mode (nonstop mode) in order to use the xCAD Client. You cannot use the xCAD Client only. Thermo Scientific GenPure...
To wall mount the xCAD Client wall version unscrew the 4 screws (see red arrows in the xCAD picture) on the bottom of the xCAD Client and Client remove the wall mount bracket. Screws Wall bracket GenPure Thermo Scientific...
Page 49
Attach the wall mount bracket to the wall by screwing in the 4 supplied screws with a philips screw driver into the wall where you have put in Screw the plugs before. Philips screw- driver Wall mount bracket Thermo Scientific GenPure...
Page 50
Server When the xCAD Client is finish connected to the xCAD Server the xCAD Server must be in operating mode (nonstop mode) in order to use the xCAD Client. You cannot use the xCAD Client only. GenPure Thermo Scientific...
Page 51
Wall mount bracket wall. Screw in the 4 screws (see red arrows) which you unscrewed in step 2 to attach the xCAD Client xCAD on the wall mount bracket. Client Wall mount bracket Screws Cross screw driver Thermo Scientific GenPure...
Screw hooks Dowels Lift the System and hang the back side of it onto the screw hooks. Screw hooks in wall Lifting and carrying the system should be completed by 2 people. Backside Wall system GenPure Thermo Scientific...
Page 53
Installation Wall mounting GenPure xCAD Plus system Figure 1. Holes for wall mounting Thermo Scientific GenPure...
The removable line cord must be shown to the Universal adapter bottom. Plug the connecting cable (appliance cable) in the power pack socket. Power pack Connecting cable Do not bring the power pack in contact with water. Risk of an electrical shock. GenPure Thermo Scientific...
Page 55
Connect the power pack to the ultrapure water system (48V 4-pin power supply connector, connector 8) and to an earthed 100 - 250V, Power 50/60Hz socket. supply connector Switch the system on. The system is now ready for use. Thermo Scientific GenPure...
Connect the hose which has a R3/4 female nut (1) from the raw water tap to the R3/4” input of the ion exchanger. Make connection from the R3/4 output of the ion exchanger to the feedwater connector of the GenPure system by using the hose (2) that is contained in the assembly kit. GenPure Thermo Scientific...
The TOC value is calculated by taking the difference between the values measured by the measuring cells QIA300 and QI302. The measurement range is 0 - 30 ppb. When this range is exceeded, the number 99 is shown in the display instead of the measured value. In Stand-by operation, “___” is shown. Thermo Scientific GenPure...
Page 60
How the system functions GenPure Thermo Scientific...
0.2 μm final filter, open the white knurled screw. 0.2 μm final filter Knurled screw Do not close the knurled screw until pure water runs out of the opening at the knurled screw continuously. Rinse about 500 ml of water through the final filter. GenPure Thermo Scientific...
• UV lamp intensity and operating time Also enables values to be • Cartridge number and maintenance interval changed in the menu mode. • Rinsing • Disinfection • System errors display • System errors print-out • Unlocking the factory setting with your password. GenPure Thermo Scientific...
Operating elements xCAD Server and xCAD Client Flow chart of menu system control xCAD Server Flow chart of menu system control xCAD Server SW version 1.1 Thermo Scientific GenPure...
• Setting for the feedwater resistance values in the menu. operation of the UV • UV lamp intensity and operating time lamp. Also enables • Unlocking the factory setting with your values to be changed in password the menu mode. GenPure Thermo Scientific...
Operating elements xCAD Server and xCAD Client Flow chart control unit xCAD Client Flow chart control unit xCAD Client SW version 1.1 Thermo Scientific GenPure...
Page 68
Operating elements xCAD Server and xCAD Client Flow chart control unit xCAD Client GenPure Thermo Scientific...
(measuring cell LF1) and temperature are also displayed. The displays of messages and measured values are independent of the operating mode. The TOC value is not shown in Interval mode. The display shows: GenPure Thermo Scientific...
The standstill time is given by the difference between half an hour and the interval pump time, so that the pump and the solenoid valve are actuated in a half-hourly rhythm. The TOC value is not shown in this operating mode. The display shows: Thermo Scientific GenPure...
The operating time of the UV-lamp is recorded and the “UV duration” error message is brought to display when the limiting value set for this time is exceeded. TOC measurement is also carried out during the time that the UV-lamp is glowing only. The display shows: GenPure Thermo Scientific...
Volumetric dispense is supported in all versions. The display shows: OFF mode A second press on the ON/Off-button causes the display to go dark and all text output on the display to be extinguished. No outputs are now switched. Thermo Scientific GenPure...
Set the limiting value using the arrow buttons (see Settings with the arrow buttons). With settings above 50 μS/cm, the limiting value is switched off and the word off appears in the display. Press the Menu-button once then the display shows: GenPure Thermo Scientific...
Limiting value setting range: 0.055 - 5.000 μS/cm Basic setting: 0.200 μS/cm With settings above 5.000 μS/cm, the limiting value is switched off and “Off ” is shown in the display.Press the menu-button twice then the display shows: Thermo Scientific GenPure...
The operating time counter for the filter cartridge is set back on entry of a valid serial number.Press the menu-button 4 times then the display shows: For more details see under section “Change the ultrapure cartridge” page GenPure Thermo Scientific...
The remaining rinsing time is counted down and displayed during rinsing. Step Action Figure Press the menu-button 5 times then the display shows: Confirm rinse by putting the enter button. The rinsing is started for 30 sec. Thermo Scientific GenPure...
Change the filter cartridge with the disinfection cartridge (see under chapter “Disinfection” page 87). Confirm with enter. The Disinfection is started for 30 min, indicating the remaining time. The completely process is described under section “Disinfection” page GenPure Thermo Scientific...
Now you can see two last saved errors with date and time. Print out of Data In this menu, the current system data can be printed via a connected printer. Press the menu-button 8 times then the display shows: Thermo Scientific GenPure...
Client display changes to the same color of the xCAD Server display. When the xCAD Client is registered the both display on the xCAD Server and xCAD Client jumps automatically to the display message “Interval or Nonstop mode”. GenPure Thermo Scientific...
The language can be changed in this menu. The choice is between English, French and German. The setting can be changed with the arrow buttons. Basic setting: English After entering the OEM menu press the menu-button once then the display shows: Thermo Scientific GenPure...
In this menu, a choice is given as to whether measured values are to be displayed in the conductivity unit or the specific electric resistance unit. Basic setting: Conductivity After entering the OEM menu press the menu-button 4 times then the display shows: GenPure Thermo Scientific...
71 you can read how to rinse the system manually. Step width: Setting range: 10 - 60 sec. Basic setting: 30 sec. After entering the OEM menu press the menu-button 7 times then the display shows: Thermo Scientific GenPure...
Specification of the voltage in % of the maximum supply voltage value. Basic setting for Interval mode: 20 % (for recirculation) Basic Setting for Nonstop mode: 45 % (for dispensing ultrapure water) After entering the OEM menu press the menu-button 10 times then the display shows: GenPure Thermo Scientific...
The real time clock can be set in this menu. Basic setting: The actual date Setting range: 1-12 Month, 1-31 Day, 0-24 h, 0-60 min. After entering the OEM menu press the menu-button 13 times then the display shows: Thermo Scientific GenPure...
In Nonstop operation, a set of data is issued to the printer once only. The interface has a transmission rate of 9600 bits/sec., 8 data bits, 1 stop bit and no parity. The SUB-D socket assignment is: PIN 2: TXD PIN 3: RXD PIN 5: GND GenPure Thermo Scientific...
UV-intensity, 0 is entered in place of measured values for these functions! Code message: Whenever a code number is entered in system control and confirmed with Enter, the code input is immediately printed out. Code identification (see “Code lock” page 82). Print-out: Thermo Scientific GenPure...
Menu + OEM menu All levels 0150 0450 0750 0250 0550 0850 0350 0650 0950 Each access via the code is printed out by the printer (RS 232) complete with date, time and the code number used. The display shows: GenPure Thermo Scientific...
Months to be replaced. systems with a UV lamp) *Please keep in mind that the life of your consumable is directly dependent on the quality of the feed water and the amount of water used daily. GenPure Thermo Scientific...
Locate the new ultrapure cartridge and insert it into the system. Plug the quick connects correctly onto the new ultrapure cartridge. You will know they are Outlet attached when an audible “click” is heard. Quick Replace the cartridge cover. connector Inlet Thermo Scientific GenPure...
Page 92
UV button to go to the next value. c. When you are finished, press enter and the new serial number is saved. You can only use a serial number one time. Discard at least 5 liters of water. GenPure Thermo Scientific...
Wear protective gloves for handling syringe of cleaning solution. Wear safety goggles when working with cleaning solution. Please observe the information given in the safety data sheet supplied with disinfectant to avoid possible health hazards. Thermo Scientific GenPure...
Page 94
Quick system. connectors Disinfectant cartridge See under chapter “Change the ultrapure cartridge” page 85 to put in the filter cartridge in to the system. Re-open the feedwater supply, switch the system on again. Feedwater supply GenPure Thermo Scientific...
Page 95
Save the disinfection cartridge for later use. See step 5 under chapter “Change the ultrapure cartridge” page Before dispensing water from the system, let water run out for approx 15 minutes. The system is then ready for use. Thermo Scientific GenPure...
When you are installing the new UF filter the flow arrow of the filter must be pointing to the bottom of the system. Reinstall the back panel, reopen the feedwater supply and switch on the system again. Feedwater supply GenPure Thermo Scientific...
We therefore recommend that clean gloves be worn. See chapter “Structure of the UV-lamp” page 91 where is seating the sealing ring to not damage it. Thermo Scientific GenPure...
Page 100
Once it is in place, re-mount the bracket holding the UV housing onto the system’s the mounting plate. Connection Sealing ring UV-lamp UV-lamp housing UV-lamp UV-lamp housing GenPure Thermo Scientific...
Page 101
UV-lamp back and save the new values by an automatically calibration. The UV-lamp must be switched on (Nonstop mode). The calibration process of the UV intensity can be take between 5 min. and 2 hours. Thermo Scientific GenPure...
Unpacking the new Final filter and screw in the filter counter clockwise (See arrow in the picture) Dispensing valve in the dispensing valve outlet (R 1/4” female outlet thread) Final filter Rinse about 3 liters through final filter before use. GenPure Thermo Scientific...
96). If you are trying to dispense water and nothing is coming out from the outlet, the final filter is blocked. Please look then in chapter “Trouble shooting” page 101 or change with a new one. Thermo Scientific GenPure...
Page 104
Maintenance Change and autoclave the Final filter GenPure Thermo Scientific...
Waste disposal Before you are shipping your Thermo Scientific Barnstead Ultrapure Water System to the waste disposal, only a specially trained personal is to be taken out the system of operation. When the packaging is no longer needed it can be disposed of as household waste.
• Contact Thermo for service • Faulty Dispense button Water flows out • Leaky hose connection • Check and seal the hose connection • Feedwater pressure > 6 bar • Install a pressure reducer • Contact Thermo for service Thermo Scientific GenPure...
Page 108
• Measuring cell cable break • Replace the measuring cell “Measuring cell LF1” • System control defect • Replace system control • Conductivity of ultrapure • see “Resistance < 18.2 water outside of the M ·cm” page 101 measuring range GenPure Thermo Scientific...
Page 109
• Replace the measuring cell “Temp. meas. cell.” cable • System control defect • Replace the system control Error message: • Operating hours of the filter • Replace it with a new one “change cartridge” cartridge has expired Thermo Scientific GenPure...
Replacement parts GenPure Thermo Scientific GenPure...
Page 112
Check valve 50150598 UV-Intensity sensor (only UV - TOC) 16.0222 Leakage sensor 16.0389 G fuse holder 5 x 20 mm 50143154 G fuse, 5 x 20 mm, 4.0 A 50150714 Table top power pack (not shown) 50149597 GenPure Thermo Scientific...
We ask for your understanding that our guarantee for this system is invalid when replacement parts, accessories or consumables materials from other manufactures are used in or for the system, as we have no influence on their composition or quality. Thermo Scientific GenPure...
Page 114
Replacement parts GenPure GenPure Thermo Scientific...
[l/h] Any false entry is considered to be a falsification of documents. The following point should be observed for maintenance of the quality of the system: • 1x / Weekly, acquire measured values. Thermo Scientific GenPure...
Page 122
Maintenance records GenPure Thermo Scientific...
Contact Information Thermo Scientific The address to contact when your system requires service: Overview of Thermo Scientific International Sales Organization Postal address USA: Thermo Scientific 275 Aiken Road Asheville, NC 28804 Enquiries from USA/Canada +1 866 984 3766 Sales: +1 800 438 4851...
Page 124
Contact Information Thermo Scientific Finland/Nordic/Baltic countries +358 9 329 100 Sales: +358 9 329 100 Service: France +33 2 2803 2180 Sales: +33 825 800 119 Service : Germany: Postal Address Germany: Thermo Electron LED GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold...
Flow charts 51 Consumable materials 109 Contact information 117 GenPure xCAD Plus bench version 14 Contact Information Thermo Scientific 117 GenPure xCAD Plus wall version 15 Contents 1 Holes for wall mounting 47 Data transmission via the RS 232 interface 80...
Page 126
Registering the xCAD Client to the xCAD Server 74 Registering the station 75 Replacement parts 105 Rinsing the ultrafilter 71 Return shipment 6 Ultrapure water limiting value 69 Rinsing the ultrafilter 71 UV Intensity and operating time 69 Rinsing time 77 GenPure Thermo Scientific...
Page 127
Index: UV-Lamp 66 UV-lamp Change 93 Structure 91 UV-Lamp operating time and intensity 70 Wall mounting Holes 46 Wall mounting GenPure xCAD Plus system 46 Warranty iv Waste disposal 99 Thermo Scientific GenPure...