Setting the limiting value for conductivity

Step
Action
1
Press the "Menu" button twice to enable the
limiting value for the conductivity to be set
(red LED lights up).
2
Use the "up/quit" and "down" button to
increase or reduce the value. With the "Enter"
button you switch to the next digit. After
finishing adjustment press "Enter" to confirm
the value.
3
Press the "Menu" button twice to return to
the operating mode.
Thermo Scientific
• The limiting value can only be set in μS/cm and is factory adjusted to a value of
0.1 μS/cm.
• Setting range: 0.055 - 30 μS/cm, basic setting: 0.1 μS/cm
Picture
Display
Second line
Display shows
Limit value Cond.
or
0.10
12
System control
Display shows
0.10
Temp.
Status Sensor
Display shows
0.10
Smart2Pure
59