Mounting the ball valve to the Smart2Pure system

Step
Action
1
Switch the system off. (unplug the main power
supply)
2
a. Locate the ball valve, the t-adapter and the
outer diameter 1/4" tube.
b. Remove the tube jumper from the
connection panel of the system.
c. Cut from the outer diameter 1/4" tube two
tubes with a length of 80 mm/ 3.2 inch and
attach the tubes into the t-adapter and the
connection panel. Attach the ball valve with
a third outer diameter 1/4" tube to the
t-adapter.
d. Additionally attach a fourth outer diameter
1/4"tube to the other connection of the ball
valve.
You can choose the length of the outer diameter
1/4" tubes beginning from the t-adapter for the ball
valve by yourself.
3
Switch the system on. (Connect main power supply)
Thermo Scientific
Delivered with the assembly kit is a ball valve with a t-connector to have the
opportunity to dispensed pure water ASTM Type II quality from the internal or
external tank from the Smart2Pure system.
Proceed as follows to do this action.
Connections of the Smart2Pure 12 system
Figure
b.)
Tube jumper
Dispenser back flow
c.)
Connection panel
Outer dia-
meter 1/4"
tube
d.)
Closed position off ball valve
7
Installation
Dispenser
flow
T-adapter
Outer
meter 1/4"
tube
Ball valve
Outer dia-
meter 1/4"
Ball valve
tube
Smart2Pure
dia-
45