Thermo Scientific Smart2Pure 3 Operating Instructions Manual

Thermo Scientific Smart2Pure 3 Operating Instructions Manual

Barnstead smart2pure series ultrapure water system
Table of Contents

Advertisement

Thermo Scientific
Barnstead Smart2Pure
Ultrapure water system
Operating Instructions
50148311
Revision B June 2016

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Smart2Pure 3

  • Page 1 Thermo Scientific Barnstead Smart2Pure Ultrapure water system Operating Instructions 50148311 Revision B June 2016...
  • Page 2 © 2016 Thermo Fisher Scientific Inc. All rights reserved. Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
  • Page 3: Preface

    This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents. Thermo Scientific Smart2Pure...
  • Page 4: Legal Information

    Please consult your local sales representative for details. Warranty Thermo Electron LED GmbH warrants the operational safety and functions of the Thermo Scientific Barnstead Ultrapure Water Systems only under the condition that: • The system is installed and operated as per the operation manual. Do not use this product for anything other than its intended use.
  • Page 5 This information is valid for the system that is received. For quick and correct service, please include the following information on all inquiries and replacement parts orders which relate to your system: • the serial number (located on the right side of the system on the nameplate) • the catalog number Thermo Scientific Smart2Pure...
  • Page 6: Standards And Directives

    Preface Standards and directives The Thermo Scientific Barnstead Ultrapure Water Systems complies with the following standards and directives: • Low Voltage Directive 2014/35/EC • EMC Directive 2014/30/EU • ASTM D1193-6 • RoHs 2011/65/EU Additionally, the ultrapure water system is in compliance with many other international standards, regulations and directives not listed here.
  • Page 7: Table Of Contents

    Extend of delivery ..........9 Extend of delivery Smart2Pure 3/6 ....... . . 10 Extend of Delivery Smart2Pure 12 .
  • Page 8 Replacement parts Smart2Pure 12........84 Replacement parts pure water tank for Smart2Pure 12 ....86 Smart2Pure Thermo Scientific...
  • Page 9 Index ............95 Thermo Scientific...
  • Page 10 Contents Smart2Pure Thermo Scientific viii...
  • Page 11: Chapter 1 Transport And Packaging

    Content • “Examination on receipt” page 5 • “Complaints” page 5 • “Packaging for return shipment” page 5 Do not pull the plastic foil over your head. Risk of suffocation. Use the plastic foil only for packaging. Thermo Scientific Smart2Pure...
  • Page 12: Examination On Receipt

    Ultrapure water systems are carefully inspected and packed prior to shipping, but damage could still possibly occur during transport. Lifting and carrying the Thermo Scientific Barnstead Smart2Pure ultrapure water system, e.g. to the installation location, should be carried out by two people.
  • Page 13 • Prior to send back a operated device, empty the water and dry the system and take out the cartridges. • pack the ultrapure cartridges into a bubble wrap and/or packaging foam and take it with in into the package of the Smart2Pure ultrapure water system. Thermo Scientific Smart2Pure...
  • Page 14 Transport and packaging Smart2Pure Thermo Scientific...
  • Page 15: Chapter 2 Safety Precautions

    Safety Precautions Observe these safety precautions for your own safety! The Thermo Scientific Barnstead Ultrapure Water Systems are modern water purification system intended solely for the treatment of potable water. The water it produces is not fit for drinking. Work may only be performed on the system electronics when the system has been switched off and when ESD protection is in place.
  • Page 16 • The surface or wall on which the system is to be installed or mounted must have an adequate load-carrying capacity (check the capacity and stability of the wall). The dry weight of the system is given under 'Dimensions and weight Smart2Pure 3/6' page 18 'Dimensions and weight...
  • Page 17 • Do not use any oxidative cleaning agents for cleaning the system. These can damage the system. • Proceed as follows when the system has a defect: • Switch the system off and unplug the system from power outlet. • Shut off the feedwater supply. • Contact your local service organization. Thermo Scientific Smart2Pure...
  • Page 18 Safety Precautions Smart2Pure Thermo Scientific...
  • Page 19: Chapter 3 Extend Of Delivery

    Extend of delivery Contents • "Extend of delivery Smart2Pure 3/6" page 10 • "Extend of Delivery Smart2Pure 12" page 11 • "Available Smart2Pure 3/6 systems" page 12 • "Available storage tanks for Smart2Pure 12 systems" page 13 Thermo Scientific Smart2Pure...
  • Page 20: Extend Of Delivery Smart2Pure 3/6

    Extend of delivery Extend of delivery Smart2Pure 3/6 The following items are included with the Smart2Pure 3/6 pure water system: 1x Smart2Pure 3/6 ultrapure water system: Item No.: 5012xxxx 1 x Ultrapure cartridge Item No.: 09.1020 1x RO/pretreatment cartridge for a 3 l/h system Item No.: 09.2003...
  • Page 21: Extend Of Delivery Smart2Pure 12

    Item No.: 50132215 1 x Plug angle connector 1/4“ Item No.: 14.0361 1 x T-connector 1/4“ Item No.: 14.0362 2 x Nylon plugs S6 Item No.: 21.002 2 x Countersunk wood screw 4 x 35 Item No.: 21.0069 Thermo Scientific Smart2Pure...
  • Page 22: Available Smart2Pure 3/6 Systems

    Extend of delivery Available Smart2Pure 3/6 systems Item No.: System 50129869 Smart2pure 3 Standard 50129870 Smart2pure 3 UF 50129872 Smart2pure 3 UV 50129688 Smart2pure 3 UV/UF 50129873 Smart2pure 6 Standard 50129874 Smart2pure 6 UF 50129885 Smart2pure 6 UV 50129887 Smart2pure 6 UV/UF...
  • Page 23: Available Smart2Pure 12 Systems

    Smart2pure 12 Standard 50129889 Smart2pure 12 UF 50129890 Smart2pure 12 UV 50129845 Smart2pure 12 UV/UF Available storage tanks for Smart2Pure 12 systems Item No.: System Capacity 06.5040 Pure water tank 30 liter 06.5070 Pure water tank 60 liter Thermo Scientific Smart2Pure...
  • Page 24 Extend of delivery Smart2Pure Thermo Scientific...
  • Page 25: Intended Use Of The Device

    The system allows the purification of water into the water categories mentioned in the standards of ASTM 11.01 and ASTM 11.02. The Thermo Scientific Barnstead Ultrapure Water System are designed to be installed and use in the following application areas: •...
  • Page 26 Intended Use of the device Smart2Pure Thermo Scientific...
  • Page 27: Chapter 5 Technical Specifications

    0.067-1.0 Resistance (reference temperature +25°C) M ·cm 15-10 15-10 15-10 15-10 Retention rate for bacteria and particles Tank capacity Smart2Pure 3/6 Liter External tank capacity Smart2Pure 12 Liter up to 60 up to 60 up to 60 up to 60...
  • Page 28 Permeate and concentrate performance for Smart2Pure Permeate Concentrate System All Smart2Pure 3 All Smart2Pure 6 All Smart2Pure 12 Dimensions and weight Smart2Pure 3/6 Height 545 mm / 21.46 inch Width 305 mm / 12.01 inch Depth 400 mm / 15.75 inch Base area 277 x 372 mm /10.91 x 14.65 inch...
  • Page 29 305 mm / 12.01 inch Depth 300 mm / 11.81 inch Base area 277 x 372 mm /10.91 x 14.65 inch Weight including ultrapure cartridge and 28 kg / 61.74 lbs RO/pretreatment cartridge 305 (12.01) 300 (11.81) Thermo Scientific Smart2Pure...
  • Page 30 Tube 1/4“ outer diameter Tank Flow tubing (only Smart2Pure 12) Tube 1/4“ outer diameter Tank back flow tubing (only Smart2Pure 12) Tank overflow tubing (only Smart2Pure 3/6) Tube 1/4“ outer diameter Dispenser back flow tubing Tube 1/4“ outer diameter Dispenser flow tubing Tube jumper Tube 1/4“...
  • Page 31 Space on left and right from the side of the system at least 300 mm / 11.81 inch Space to the back of the system at least 200 mm / 7.87 inch Top space at least 400 mm /15.75 inch Space to front of system Free accessibility Thermo Scientific Smart2Pure...
  • Page 32 60364-4-443) Note: The rated level of transient overvoltage is the withstand impulse voltage acc. to overvoltage category II of IEC 60364-4-443. Ventilation requirements no special requirements no special requirements Degree of pollution Airborne sound emission Sound pressure level 49 db(A) Smart2Pure Thermo Scientific...
  • Page 33: The Installation Area

    When no floor drain is available, install a water watcher to protect against water damage (available only for Europe). • A check valve is recommended in the feedwater line to prevent back flow of feedwater from the water system. Thermo Scientific Smart2Pure...
  • Page 34 • Ample working space must be provided around the system for easy and pleasant replacement of wear and replacement parts and for ease of operation (see 'Accessibility to Smart2Pure systems and pure water tank' page 21). • Easy access for operation and control of the system. Smart2Pure Thermo Scientific...
  • Page 35: Installation

    Installation Content • “Connections of the Smart2Pure 3/6 system” page 26 • “Connections of the Smart2Pure 12 system” page 28 • “Connections of the pure water tank Smart2Pure 12” page 30 • “Bring your Smart2Pure system into operation” page 31 •...
  • Page 36: Connections Of The Smart2Pure 3/6 System

    Installation Connections of the Smart2Pure 3/6 system Smart2Pure Thermo Scientific...
  • Page 37 12. 1/4 thread connector for sterile vent filter When you are remove the tube jumper between Pos. 4 and Pos. 5, you have the possibility to connect additionally the accessory hand dispenser, Item No.: 50138221(purchased separately). This hand dispenser comes with an installation guide. Thermo Scientific Smart2Pure...
  • Page 38: Connections Of The Smart2Pure 12 System

    Installation Connections of the Smart2Pure 12 system Connections of the Smart2Pure 12 system Smart2Pure Thermo Scientific...
  • Page 39 12. Quick connectors for ultrapure cartridge When you remove the tube jumper between Pos. 6 and Pos. 7, you have the possibility to connect additionally the accessory hand dispenser, Item No.: 50138221(purchased separately). This hand dispenser comes with an installation guide. Thermo Scientific Smart2Pure...
  • Page 40: Connections Of The Pure Water Tank Smart2Pure 12

    30L pure water tank 13. Tank flow connector for outer diameter 1/4“ tube 14. Tank back flow connector for outer diameter 1/4“ tube 15. Level control connector 16. Connector for optional sterile overflow 17. Connector for optional sterile vent filter Smart2Pure Thermo Scientific...
  • Page 41: Bring Your Smart2Pure System Into Operation

    Ultrapure The quick connectors are attached to the system in such cartridge a manner so as to prevent installing the ultrapure cartridge incorrectly. Thermo Scientific Smart2Pure...
  • Page 42 1/4“ tube to the feedwater connection of the Smart2Pure panel. b. After this screw in the R3/4“ union nut with the 1/4“ tube Fitting Outer diameter dirt screen onto the feedwater tap. Feedwater connection Smart2pure 3/6 Feedwater connection Smart2pure 12 Smart2Pure Thermo Scientific...
  • Page 43 1/4“ tubing to the concentrate and a second one to the tank overflow outlet. (see “Connections of the Smart2Pure 3/6 system” page b. Route both AD 1/4“ tubes with a free gravity fall to the drain. (see “Illustration of drain”...
  • Page 44: Connecting An External Tank To The Smart2Pure 12 System

    Plug in the AD 1/4“ tube to connect the tank flow connector on the pure water tank to the tank flow Tank flow Tank flow connector on the Smart2Pure 12 panel. connection tank connection Smart2pure 12 Smart2Pure Thermo Scientific...
  • Page 45: Illustration Of Drain

    If a standard drain siphon is in place, the ends of the tubes must be located at least 20 mm/ 0.75 inch above the Minimum distance drain. Attach the tubes in such a manner that they remain 20 mm/0.75 inch in their position. Drain Thermo Scientific Smart2Pure...
  • Page 46: Attaching The Wall Holder For 30 Or 60L Pure Water Tank

    60L pure water tank” page 20). • Lifting and carried out only the 30L or 60L tank by two person. • Do not lift up the 30L or 60L tank with full water capacity. Risk of ergonomic postural deformities. Smart2Pure Thermo Scientific...
  • Page 47: Attaching Wall Holder For 30L Pure Water Tank

    Place the 30L pure water tank onto the mounted wall holder bottom part with two people and then screw in the four supplied M8 hexagon nuts onto the threaded bolts from the wall holder bottom part. (see pictures next side) Thermo Scientific Smart2Pure...
  • Page 48 Installation Connections of the Smart2Pure 12 system Openings for screw on the hexagon nuts onto the wall holder bottom part. Threaded bolts Hexagon nuts Wall holder Wall holder bottom part bottom part Threaded bolts Smart2Pure Thermo Scientific...
  • Page 49 Screws Wall holder upper part After mounting the complete wall holder connect the tubing and level control cable as described at chapter “Connecting an external tank to the Smart2Pure 12 system” page Thermo Scientific Smart2Pure...
  • Page 50: Attaching Wall Holder For 60L Pure Water Tank

    Then use a 8 mm or 0.31 inch twist drill to make the holes and put in the six S8 plugs which are supplied with the wall holder. Mark up here the position for bore- holes Wall Smart2Pure Thermo Scientific...
  • Page 51 Phillips screw driver. Screws Wall Place the 60L pure water tank with two people on the wall holder and repeat the steps 4, 5 and 6 in section “Attaching wall holder for 30L pure water tank” page Thermo Scientific Smart2Pure...
  • Page 52: Wall Mounting

    Installation Connections of the Smart2Pure 12 system Wall mounting You can also mount your Smart2Pure 3/6 or Smart2Pure 12 system on the wall. To do this, use the wall-mounting fixture included in the accessories (purchased separately Item No.: 09.2212). Before you begin mounting the system on the wall...
  • Page 53: Mounting The Power Pack (Voltage Supply)

    Universal holder Universal foil Universal adapter Mains adapter Attach universal adapter to a smooth wall surface with Universal adapter Wall surface supplied screws or with glue (not provided). Thermo Scientific Smart2Pure...
  • Page 54 Power cable Now, connect the power supply unit to the 'Power supply' connection on the right side of the Smart2Pure 3/6 or Smart2Pure 12 system. Next plug the power supply to a grounded 100 - 240V, 50/60 Hz power outlet.
  • Page 55: Mounting The Ball Valve To The Smart2Pure System

    Outer dia- 1/4“ tubes beginning from the t-adapter for the ball meter 1/4“ valve by yourself. tube Closed position off ball valve Outer dia- meter 1/4“ Ball valve tube Switch the system on. (Connect main power supply) Thermo Scientific Smart2Pure...
  • Page 56 Installation Connections of the Smart2Pure 12 system Smart2Pure Thermo Scientific...
  • Page 57: Chapter 8 Flow Chart

    ASTM Type 2 system port Pure water ASTM Type 1 system Pure water Storage reservoir: Internal for 3 and 6L/h versions. dispenser External for 12L/h versions with 0.2 μm Ultrapure water filter Purification flow Recirculation flow Waste water Thermo Scientific Smart2Pure...
  • Page 58 Flow chart Smart2Pure Thermo Scientific...
  • Page 59: How The Smart2Pure System Functions

    2 to ensure the constant quality of pure water ASTM Type II. A level switch monitors the water level inside the tank. ASTM Type II water can be dispensed from the side outlets of internal tank from the connection panel or from the front valve of the external tank. Thermo Scientific Smart2Pure...
  • Page 60 How the Smart2Pure system functions Smart2Pure Thermo Scientific...
  • Page 61: Chapter 10 Initial Start Up

    Initial start up Contents • “Putting the system into operation” page 52 • “Dispensing water from the dispensing valve” page 54 • “Venting the 0.2 μm final filter” page 54 Thermo Scientific Smart2Pure...
  • Page 62: Putting The System Into Operation

    7“Installation” page 25 or shown in the bottom diagram. Connections Connections Smart2Pure 3/6 Smart2Pure 12 1. Concentrate connector for outer diameter 1/4“ tube 2. Feedwater connector for outer diameter 1/4“ tube 3. Tank overflow connector for outer diameter 1/4“ tube 4.
  • Page 63 5 minutes. After the brief self test the system begins to produce pure and ultrapure water. • If the tank content of the Smart2Pure 3/6 system is empty it is possible that air is coming from the tank into the ultrafilter (only systems with ultrafilter). Thereby you can not dispense water from the sampling tap on the system.
  • Page 64: Dispensing Water From The Dispensing Valve

    0.2 μm final filter, open the white knurled screw. 0.2 μm final filter Knurled screw Do not close the knurled screw until pure water runs out of the opening at the knurled screw continuously. Rinse about 500 ml of water through the final filter. Smart2Pure Thermo Scientific...
  • Page 65: Chapter 11 Operating Elements

    D = Menu button: switches the Menu to the next submenu. E = Down button: decreases adjustable values. F = up/Quit button: increases adjustable values, quits acoustic alarms. G = Enter button: confirms changes, allows to adjust the cursor. Thermo Scientific Smart2Pure...
  • Page 66 Operating Elements Smart2Pure Thermo Scientific...
  • Page 67: Chapter 12 System Control

    58 • “Setting the limiting value for conductivity” page 59 • “Setting the limiting value for temperature” page 60 • “Error message and acoustic error signal” page 61 • “Communication” page 61 • “Potential-free contact” page 62 Thermo Scientific Smart2Pure...
  • Page 68: Menu

    μS/cm °C M ·cm Use the “up/Quit“ and “down“ button to choose the resistance value for measuring Display shows mode (μS/cm or M ·cm) and than choose the mono measurement mode “c“ or dual measurement mode “ct“. Smart2Pure Thermo Scientific...
  • Page 69: Setting The Limiting Value For Conductivity

    With the “Enter“ Display shows button you switch to the next digit. After finishing adjustment press “Enter“ to confirm the value. 0.10 Display shows 0.10 Press the “Menu“ button twice to return to the operating mode. Thermo Scientific Smart2Pure...
  • Page 70: Setting The Limiting Value For Temperature

    With the “Enter“ Display shows button you switch to the next digit. After finishing adjustment press “Enter“ to confirm the value. 35°C Display shows 35°C Press the „“Menu“ button once to return to the operating mode. Smart2Pure Thermo Scientific...
  • Page 71: Error Message And Acoustic Error Signal

    Automatically the conductivity is scaled to three significant places. Output is made 1x per hour. When the system is in operation mode , a press on the „“down“ button triggers a measured value output at the serial interface. Print out example: 18.2 M ·cm, 23.4°C Thermo Scientific Smart2Pure...
  • Page 72: Potential-Free Contact

    System control Potential-free contact The system is equipped on the systems panel with a potential free contact for control external equipment. The maximum connected load is 30V, 2A The Pin assignment of the 5 pin socket is: Smart2Pure Thermo Scientific...
  • Page 73: Chapter 13 Maintenance

    68 • "Changing the ultrafilter" page 63 • "UV-reactor assembly" page 73 • "Replacing the UV lamp" page 74 • "Replacing the 0.2 μm final filter" page 76 • "Autoclaving the 0.2 μm final filter" page 76 Thermo Scientific Smart2Pure...
  • Page 74: Maintenance Intervals

    Or when the ultrapure cartridge has cartridge 09.2006 shorter life time than excepted. 09.2012 Please note that the lifetime of the wear parts is a direct function of the quality of the feedwater and the daily volume of water that is used. Smart2Pure Thermo Scientific...
  • Page 75: Replacing The Ultrapure Cartridge

    When you hear an audible click you can be sure that the quick connectors have been inserted correctly. The quick-connectors are attached to the system in such a ultrapure manner so it is to prevent any confusion (switching). cartridge Thermo Scientific Smart2Pure...
  • Page 76: Replacing The Ro/Pretreatment

    Press on the three quick connectors on the cover of the Quick RO/pretreatment cartridge and remove the used connectors cartridge from the system. RO/pretreat- ment cartridge Locate the new RO/pretreatment cartridge and insert the cartridge into the system. Smart2Pure Thermo Scientific...
  • Page 77 The quick connectors are attached to the system in such ment cartridge a manner so as to prevent any confusion switching. Replace the cover over the RO/pretreatment cartridge and switch the system on (connect main power supply). Thermo Scientific Smart2Pure...
  • Page 78: Cleaning

    • Depending on the version of your Smart2Pure system (with internal 5L tank Smart2Pure 3/6 or Smart2Pure 12 with external tank), a complete cleaning will take between 1 to 5 hours. Required for cleaning is a cleaning adapter (item no: 50133431).
  • Page 79 There is no need to worry should you notice a chlorine smell during the entire cleaning process, as there is no risk of the limit for chlorine gas in closed rooms being exceeded. Let the Smart2Pure system run for 1 hour. Thermo Scientific Smart2Pure...
  • Page 80 (see "Replacing the ultrapure cartridge" page d. Switch the system on (connect the main power Dispensing valve external tank supply) and drain the first 10 L through the dispensing valve. Quick connectors Adapter for cleaning Smart2Pure Thermo Scientific...
  • Page 81: Changing The Ultrafilter

    Ensure that the white O-rings on the 1/4“ inch tubes are not lost in the process. You will need these again for Fitting re-attaching the tube. Measuring cell Ultrafilter Threaded connection fitting Thermo Scientific Smart2Pure...
  • Page 82 UV assembly into place (see Step 2c). Mounting a. Replace the cover on the system, open the bracket feedwater tap and switch the system on (connect main power supply). Dispense the first 5 L of Ultra-filter water. UV lamp Smart2Pure Thermo Scientific...
  • Page 83: Uv-Reactor Assembly

    Item No.: 09.1002 UV lamp O-ring Item No.: 21.5030 Immersion tube Immersion tube Item No.: 22.0040 UV lamp casing Item No.: 22.0092 Stopper Item No.: 14.0187 These stoppers are located in the UV lamp casing Stopper Item No.: 21.5008 Thermo Scientific Smart2Pure...
  • Page 84: Replacing The Uv Lamp

    UV lamp out by about 1cm while still plugged in. a. Carefully unplug the plug for the UV lamp. b. Now, carefully pull the UV lamp up while turning it UV-lamp slightly in a clockwise direction at the same time. lamp casing Smart2Pure Thermo Scientific...
  • Page 85 You can then re-attach the plug to the UV lamp and retighten the UV assembly on the mounting plate using the two retaining brackets and the two screws that you removed previously. UV lamp lamp casing lamp UV lamp casing Thermo Scientific Smart2Pure...
  • Page 86: Replacing The 0.2 Μm Final Filter

    (see "Replacing the 0.2 μm final clogged. In this case, read the description in section filter" page 76). "Trouble shooting" page 79, or replace the 0.2 μm final filter with a new one. Smart2Pure Thermo Scientific...
  • Page 87: Chapter 14 Waste Disposal

    Only specially trained personal can take the system out of operation and dispose it properly. If you have a used or broken UV-lamp, contact your local Thermo Scientific service organization or waste disposal company.
  • Page 88 Waste disposal Smart2Pure Thermo Scientific...
  • Page 89: Chapter 15 Trouble Shooting

    • Wait until the tank is again of empty tank filled and a rinse is triggered Control panel non responsive • Microprocessor locked up • Unplug the mains plug for 5 seconds. • PCB error • Contact the Thermo Fisher Scientific service department Thermo Scientific Smart2Pure...
  • Page 90 'Red LED in second line is light ductivity measuring cell nection up “Status Sens.“ • Measuring cell defect • Contact Thermo Scientific for replacing Error message: • Interruption in cable to con- • Check the cable and plug con- Red LED in second line is light...
  • Page 91: Chapter 16 Replacement Parts

    Please note that the use of spare parts, accessories or wear parts from other manufacturers will nullify the warranty for this unit. Contents • “Replacement parts Smart 2Pure 3/6” page 82 • “Replacement parts Smart2Pure 12” page 84 • “Replacement parts pure water tank for Smart2Pure 12” page 86 Thermo Scientific Smart2Pure...
  • Page 92: Replacement Parts Smart 2Pure 3/6

    Replacement parts Replacement parts Smart 2Pure 3/6 Smart2Pure Thermo Scientific...
  • Page 93 Replacement parts Pos. Designation Article number Ultra pure water tank 17.0136 RO/pretreatment cartridge with activated carbon and 09.2003 for Smart2Pure 3 09.2006 RO/pretreatment cartridge with activated carbon and for Smart2Pure 6 Ultrapure cartridge 09.1020 Final 0.2 micron filter 09.1003 Sterile vent filter 22.0091...
  • Page 94: Replacement Parts Smart2Pure 12

    Replacement parts Replacement parts Smart2Pure 12 Smart2Pure Thermo Scientific...
  • Page 95 Recirculation solenoid valve 15.0062 Check valve 50150598 Check valve 15.0114 Control Board with display 16.0351 Interface 16.0337 Fuse holder for glass tube fuse 50143154 Glass tube fuse 5x20mm, 3,15 A, slow fuse 50150714 Tabletop power pack (not shown) 50149597 Thermo Scientific Smart2Pure...
  • Page 96: Replacement Parts Pure Water Tank For Smart2Pure 12

    Pos. Designation Article number Dispensing valve 14.0250 Pure water tank 30 liters 18.0114 Pure water tank 60 liters 18.0115 Float switch 50133991 Sterile tank overflow (option) 50148041 Replacement UV-lamp (option) 09.5002 Sterile vent air filter (option) 06.5003 Smart2Pure Thermo Scientific...
  • Page 97: Chapter 17 Consumables

    Consumables Designation Item No. Ultrapure cartridge 09.1020 UV lamp 09.1002 Ultrafiltration module 50133981 Sterile filter 0.2 μm 09.1003 RO/pretreatment cartridge for: Smart2Pure 3 09.2003 Smart2Pure 6 09.2006 Smart2Pure 12 09.2012 Sterile vent air filter 22.0091 vent filter 06.5002 Accessories Designation Item No.
  • Page 98 Consumables Smart2Pure Thermo Scientific...
  • Page 99: Chapter 18 Terminal Assignment

    Terminal assignment Thermo Scientific Smart2Pure...
  • Page 100 Terminal assignment Smart2Pure Thermo Scientific...
  • Page 101: Chapter 19 Maintenance Record

    Last change water volume Last of ultrapure Last check of concentrate flow Signature flow cleaning cartridge [l/h] The following points must be observed in order to ensure the quality of the system. 1x / Weekly, enter measured values. Thermo Scientific Smart2Pure...
  • Page 102 Maintenance record Smart2Pure Thermo Scientific...
  • Page 103: Chapter 20 Contact Information Thermo Fisher Scientific

    Contact information Thermo Fisher Scientific Contact address for service: Overview of Thermo Scientific international sales organization Postal address in USA: Thermo Scientific 275 Aiken Road Asheville, NC 28804 Inquiries from USA/Canada Sales: +1 866 984 3766 Service: +1 800 438 4851...
  • Page 104 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Smart2Pure Thermo Scientific...
  • Page 105: Index

    34 Connections of the pure water tank Smart2Pure 12 30 Connections of the Smart2Pure 12 system 28 Legal Information ii Connections of the Smart2Pure 3/6 system 26 Consumables 87 Contact information Thermo Fisher Scientific 93 Maintenance 63 Maintenance Intervals 64...
  • Page 106 Index: R Permeate and concentrate performance for Smart2Pure Venting the sterile filter 53 Preface i Product water quality ASTM Type I (on dispensing valve) 17 Product water quality ASTM Type II (tank quality) 17 Wall mounting 42 Putting system into operation 52 Warranty ii Waste disposal 77 Water connections 20...
  • Page 107 Thermo Scientific Smart2Pure...
  • Page 108 Index:...
  • Page 109 Index Thermo Scientific Smart2Pure...
  • Page 110 Index: Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. thermoscientific.co © 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific 81 Wyman Street Waltham, MA 02451...

This manual is also suitable for:

Smart2pure 6Smart2pure 12

Table of Contents