Feedwater requirements
Source
Silt density index (SDI)
Conductivity (reference temperature +25 °C) 1200 μS/cm (0.0008 M ·cm)
Free chlorine
Manganese content
Iron content
pH range
Temperature
Pressure
Product water quality ASTM Type I (on dispensing valve)
Conductivity (reference temperature +25 °C)
Resistance (reference temperature +25°C)
TOC value
RNase
DNase
Bacteria
Bacterial endotoxins
Particles
Performance
Product water quality ASTM Type II (tank quality)
Conductivity (reference temperature +25 °C)
Resistance (reference temperature +25°C)
Retention rate for bacteria and particles
Tank capacity Smart2Pure 3/6
External tank capacity Smart2Pure 12
Thermo Scientific
Check at regular intervals the quality of your feedwater.
potable tap water pre-treated with required 09.4000 (sold separately)
Bacteria should be < 0.01
max. 0.1 ppm
max. 0.05 ppm
max. 0.05 ppm
4 - 11
+2°C - +35°C
1 - 6 bar /14 - 87 PSI (at a pressure > 6 bar / 87 PSI a pressure reducer
must be installed upstream of the system.
μS/cm
M ·cm
ppb
ng/ml
pg/ul
CFU/ml
EU/ml
μm/ml
l/min
μS/cm
M ·cm
%
Liter
Liter
Standard UV
0.055
0.055
18.2
18.2
5-10
1-5
- -
- -
- -
- -
< 0.1
< 0.1
- -
- -
< 0.2
< 0.2
1.0
1.0
Standard UV
0.067-1.0
0.067-1.0
15-10
15-10
99
99
5
5
up to 60
up to 60
5
UF
UV/UF
0.055
0.055
18.2
18.2
5-10
1-5
- -
< 0.003
- -
< 0.4
< 0.1
< 0.1
< 0.001
< 0.001
< 0.2
< 0.2
0.6
0.6
UF
UV/UF
0.067-1.0
0.067-1.0
15-10
15-10
99
99
5
5
up to 60
up to 60
Smart2Pure
17