Replacing the ultrapure cartridge

Step
Action
1
Switch the system off (unplug main power supply).
2
Remove the cover from the ultrapure cartridge by press the
snapper and pull the cover toward the front to remove it.
3
Press on the two quick connectors at the inlet and outlet of
the ultrapure cartridge and remove the used cartridge from
the system.
4
We recommend performing a cleaning when an existing
ultrapure cartridge is replaced.
5
Locate the new ultrapure cartridge and insert the cartridge
into the system.
6
Insert the quick connectors into the connecting points on
the ultrapure cartridge. When you hear an audible click you
can be sure that the quick connectors have been inserted
correctly.
The quick-connectors are attached to the system in such a
manner so it is to prevent any confusion (switching).
Thermo Scientific
Replace the ultrapure cartridge with a new one when the ultrapure water
conductivity is purity dropped below acceptable levels or the replacement interval is
due.
Figure
Snapper
Cover
Refer to section
"Cleaning"
an explanation of how to perform cleaning.
13
Maintenance
Quick
connectors
Ultrapure
cartridge
on
page 68
for
ultrapure
cartridge
Smart2Pure
65