Dräger Narkomed 2B Operator's Instruction Manual page 21

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

Fresh Gas Outlet
(Canada)
The fresh gas outlet, located on the front of the machine, delivers the
fresh gas mixture to the patient breathing system. The fresh gas
mixture consists of oxygen, nitrous oxide, optional gases, and vapors of a
liquid anesthetic.
The fresh gas outlet has a dual fitting that lets you use a gas hose with
either of the following types of fittings:
• A 15 mm male fitting, such as those supplied with North
American Dräger Absorbers and Bain Circuit Adapters. When
using a 15 mm fitting, place the spring-loaded locking bar over
the male fitting to secure it to the female fitting.
• A 22 mm female fitting with a load-bearing threaded mount, such
as those for Magill circuits or ISO-type non-rebreathing adapters.
When using an ISO-type non-rebreathing adapter, swing the
spring-loaded locking bar to the side to gain access to the
threaded load-bearing fitting.
TABLETOP
FRESH GAS
HOSE
General Description
RELEASE TO
LOCK
Section 2
O2 FLUSH CONTROL
LUER-FITTING
FOR MANUAL
SPHYGMOMANOMETER
(OPTIONAL)
HANDLE
FRESH GAS
LOCKING BAR
OP10012