Dräger Narkomed 2B Operator's Instruction Manual page 33

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

Oxygen Supply
Pressure Alarm
RED O2 SUPPLY
PRESSURE
LOW INDICATOR
YELLOW AC
POWER FAIL
INDICATOR
YELLOW
BATTERY LOW
INDICATOR
GREEN
BATTERY TEST
INDICATOR
The oxygen supply pressure alarm is activated when the oxygen supply
pressure in the system falls below approximately 37 psi. When the alarm
is activated, the red O
panel lights. In addition, the Caution message LO O
on the central alarm display and an intermittent audible alarm sounds.
Depending on the anesthesia machine's configuration, a 7-second whistle
may also sound.
NOTE:
When one source of oxygen pressure (either pipeline or
reserve cylinders) fails, but the other source is able to
maintain proper pressure within the machine's oxygen supply
lines, the oxygen supply pressure alarm is not activated.
O SUPPLY PRESSURE
2
AC
POWER FAIL
BATTERY
LOW
BATTERY TEST
ON
STANDBY
SYSTEM POWER
SUPPLY PRESSURE LED on the main switch
2
Section 2
General Description
BATTERY
TEST PUSH
BUTTON
SYSTEM
POWER
SWITCH
SUPPLY appears
2
OP20050
2-21