Monitor Controls

You use the minute volume alarm keys to set minute volume alarm
limits and to turn the volume alarms on and off. These keys are in the
bottom row of the key bank to the right of the right-hand display screen.
LOW MINUTE VOLUME
ALARM LIMIT
ADJUSTMENT KEY
TID VOL
0.58
INCREMENT
KEY
Fixed alarms are provided for low tidal volume, low minute volume, and
reverse flow through the sensor. Apnea volume alarms are generated at
15 seconds (Caution) and 30 seconds (Warning) if the respiratory volume
monitor does not sense a valid breath.
The Narkomed 2B's volume alarms are automatically enabled when the
ventilator power switch is turned to the ON position. A disconnected
sensor causes a sensor disconnect alarm.
Respiratory Volume Monitoring
BPM
MIN VOL
10
5.6
3.0
Section 5 - Operation
DECREMENT
MINUTE VOLUME ALARMS
VOLUME
ALARMS
LO
DISABLE
OP20031
DISABLE KEY
KEY
VOLUME
ALARMS
5-11-3