Installing the
For breathing pressure monitoring at the patient Y-piece, install the
Y-piece Pilot
longer pilot line (which has a quick-connect fitting on one end and a
Line
Luer type fitting on the other end) as follows:
1. Insert the quick-connect fitting on the pilot line into the quick-
2. Insert the Luer fitting on the pilot line with a Luer fitting on either
connect fitting on the breathing pressure monitor interface panel.
the patient Y-piece or a 15 mm adapter on the patient side of the
Y-piece. Use the four plastic hose clips attached to the pilot line to
mount it on either of the breathing hoses leading to the Y-piece.
PLASTIC
HOSE CLIPS
BREATHING
PRESSURE
Section 5 - Operation
Breathing Pressure Monitoring
BREATHING PRESSURE
MONITOR INTERFACE PANEL
QUICK-CONNECT
FITTING
MALE LUER-LOCK
FITTING
PILOT
LINE
OP10102
5-12-3