Section 6
Routine Maintenance and Cleaning
Manual Sphygmo-
manometer
Maintenance
6-6
SECONDARY WASTE
GAS INPUT PORT
INPUT
PORT CAP
OP00277
Under typical conditions, the only cleaning the manual
sphygmomanometer requires is a wiping down with a liquid disinfection
agent. However, if further disinfection is required, remove the
sphygmomanometer gauge assembly, hoses, and blood pressure cuff from
the anesthesia machine and sterilize them with ethylene oxide gas (cold
cycle), followed by appropriate aeration according to the sterilizer
manufacturer's instructions.
NOTE:
Do not autoclave the gauge assembly; it cannot withstand the
heat of autoclaving.
SAFETY RELIEF
VALVE HOUSING
PLASTIC
WASHER
SAFETY
RELIEF VALVE
O-RING
WASTE GAS
INPUT PORT
WASTE GAS
EXHAUST PORT