Monitor Controls - Dräger Narkomed 2B Operator's Instruction Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

Monitor Controls

Right-Hand Screen—The numerical value for inspiratory oxygen is
shown on the right-hand screen. To the right of this number, in
smaller type, are the high and low oxygen concentration alarm
limits.
OXYGEN
CONCENTRATION
VALUE
OP20057
You use the oxygen monitor control keys to set oxygen concentration
alarm limits and calibrate the oxygen sensor. These keys are in the top
row of the key bank to the right of the right-hand display screen.
LOW OXYGEN
CONCENTRATION
ALARM LIMIT KEY
OP20022
HIGH OXYGEN
CONCENTRATION
ALARM LIMIT KEY
Section 5 - Operation
Oxygen Monitoring
HIGH OXYGEN CONCENTRATION
% OXYGEN
OXYGEN ALARMS
100
40
30
HI
LO
CONCENTRATION ALARM LIMIT
% OXYGEN
OXYGEN ALARMS
100
40
HI
LO
INCREMENT
KEY
ALARM LIMIT
O
CAL
2
21%
LOW OXYGEN
DECREMENT
KEY
O2 CAL
21%
OXYGEN
CALIBRATION
KEY
5-10-3