Maintenance Intervals - Dräger Savina Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Savina:
Table of Contents

Advertisement

Care

Maintenance Intervals

Clean and disinfect equipment and/or components
before any maintenance procedures – and before returning
for repair!
O
sensors
2
Microfilter
Dust filter set
Blower unit
Lead-gel battery
Filter in O
inlet
2
Real-time clock
Pressure reducer
Equipment inspection
and service
74
replace when the following message is displayed:
!!! O
measurement inop.
2
in case of calibration is no longer possible.
Used O
sensors can be returned to
2
Dräger Medical AG & Co. KGaA for disposal.
Replace after 1 year, see page 75.
Dispose of with normal domestic waste.
Replace after 1 year, see page 76.
Dispose of with normal domestic waste.
To be replaced by trained service personnel after
20.000 hours' operation or after 8 years – whichever is earlier.
To be replaced by trained service personnel after 2 years.
Lead-gel batteries should be disposed of as special waste,
see page 78.
To be replaced by trained service personnel every 6 years.
To be replaced by trained service personnel every 6 years.
To be replaced by trained service personnel every 6 years.
Should be carried out for the first time after 2 years or
6,000 hours' operation, whichever is earlier.
Thereafter annually or after 6,000 hours' operation, whichever
is earlier.
For other servicing (e.g. six-monthly inspection, etc.), contact
DrägerService.
Only original Dräger parts should be used.