Description
Automatic volume compensation
If volume-controlled ventilation modes (IPPV, SIMV) are
applied in »Mask/NIV« mode, Savina increases the applied
tidal volume V
by the amount lost due to leakage during
T
inspiration. If a tidal volume V
of 500 mL has been set and
T
50 mL are lost during inspiration, Savina will automatically
deliver 550 mL. Because unlimited leakage compensation is
not advisable, Savina compensates for leakage up to 100 % of
the set tidal volume.