Preparation

Checklist

To be carried out every time before using the device.
Device is complete and ready for operation.
The test lung is connected.
What
Operation of lamps/
LEDs and audible alarm
Operation of audible
alarm for mains power
failure
Apnoea alarm
"PEEP high" alarm
"Airway pressure high"
alarm
"MV low" alarm
Tightness of breathing
system
When the check has been completed successfully, Savina is
ready for use.
Do not use device unless all the test steps have been
completed.
Adapt all settings and alarm limits to current patient
situation.
96
How
Switch device on:
Hold down »Alarm Reset« for about
3 seconds:
Set ventilation mode CPAP,
set PEEP 10 mbar,
steadily inflate and deflate the test lung to
simulate spontaneous breathing, then stop
simulation:
Press »Alarm Reset«.
Set ventilation mode IPPV,
set Paw
to 100 mbar,
keep flow sensor tight:
Press »Alarm Reset«.
Set pressure limit Pmax to OFF, set Paw
below the current max. airway pressure:
Set Paw
to 100 mbar.
Press »Alarm Reset«.
If necessary, set pressure limit Pmax to ON.
Set MV
higher than the current MV
measured value:
Set MV
lower than the current MV
measured value again.
Press »Alarm Reset«:
Set high flow acceleration FlowAcc, press and
hold down the »Insp. hold« key and observe
pressure curve:
What should happen
All lamps (except the lamps for power supply)
and LEDs light up, audible alarm sounds.
Audible alarm sounds.
At the end of the apnoea alarm time T
audible alarm sounds, the warning !!! Apnoea
appears on the screen.
Audible alarm sounds, red lamp flashes,
the warning !!! PEEP high appears on the
screen.
Audible alarm sounds. The warning
!!! Airway pressure high appears on
the screen.
Audible alarm sounds, the red alarm lamp
flashes, the warning !!! MV low appears on
the screen.
Plateau pressure remains constant.
,
Apnoea