Fitting The Fisher & Paykel Humidifier - Dräger Savina Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Savina:
Table of Contents

Advertisement

Preparation
Connecting the ventilating hoses
Do not use anti-static or conductive hoses
1 Hang the hinged arm from the rail on the left-hand side
and tighten screws.
Connect ventilation hoses, and note length of hoses
(in metres).
The total length of the inspiratory hoses between
Aquapor and the Y-piece must be not less than 1.1 m,
to avoid the risk of overheating the breathing gas.
2 Turn ports in direction of hoses.
3 Install water traps in vertical position at the lowest point
of their hose lines.
Connect the Y-piece, with the rubber sleeve of the Y-piece
on the inspiratory side.
Insert Y-piece into bracket of hinged arm.
Fitting the Fisher & Paykel humidifier
Prepare the humidifier in accordance with the separate
instructions for use.
1 Hang the humidifier with bracket to the humidifier holder
rail and tighten the screws.
Do not use an additional heat and moisture exchanger
(HME) at the same time as a humidifier.
Risk of increased breathing resistance because of
condensation.
*
DIN VDE 0750 Part 215:
The use of anti-static or electrically conductive material in the
breathing system of the lung ventilator is not considered to
contribute any improvement in safety. On the contrary, the use
of these materials increases the danger of electric shock to the
patient and of fire due to the presence of oxygen.
84
*
.
1
0,6m
3
0,6m
0,6m
0,6m
3
1
2
2
0,4m