Care
Care
Observe the hospital hygiene regulations.
Clean and prepare the machine after each patient.
Recommendation:
Change the hose system and expiration valve every week.
Keep the replacement systems ready.

Dismantling

Removing parts

Switch off the ventilator and breathing gas humidifier,
and remove their power plugs.
1 Drain condensate from water traps and ventilation hoses.
Drain the water trap on the expiration valve.
2 Drain the water container of the humidifier.
Ventilation hoses
Remove ventilation hoses from the device ports.
Remove the water traps from the ventilation hoses.
Remove the water containers from the water traps and
the expiration valve.
Prepare the ventilation hoses, water traps and associated
water jars, and the Y-piece for disinfection and cleaning
by autoclaving.
3 Remove the bacterial filter and treat it in accordance with
the accompanying instructions for use.
68