Ippv, Ippv Assist , Cppv - Dräger Savina Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Savina:
Table of Contents

Advertisement

IPPV, IPPV
, CPPV
Assist
Intermittent Positive Pressure Ventilation
Volume-controlled ventilation with a fixed mandatory minute
volume MV, set with the tidal volume V
For patients unable to breathe spontaneously, see page 98 for
details.
Intermittent Positive Pressure Ventilation, Assist
For patients with partial spontaneous breathing.
Continuous Positive Pressure Ventilation
Ventilation with continuous positive airway pressure.
This mode is not displayed on the screen as a ventilation mode.
Set IPPV by means of the keys for the ventilation parameters:
— Tidal volume »V
«,
T
— Inspiration time »T
«,
insp
— Frequency »f«,
— O
concentration »O
«,
2
2
— Positive end-expiratory pressure »PEEP«.
If PEEP is greater than 0 the airway pressure remains
constantly in the positive range (CPPV).
Pressure limit Pmax
The pressure peak can be avoided in IPPV and IPPV
setting a pressure limit. Thereby the PLV ventilation mode can
be reached (Pressure Limited Ventilation).
To set the pressure limit Pmax:
Set Pmax to ON, see page 51.
Set pressure limit value by means of »Pinsp« key.
When setting the pressure limit, ensure that the tidal
volume V
is still being applied, as otherwise Savina will
T
display a "!! Tidal volume low" message. Then:
Set pressure limit higher,
or
Increase inspiration time »Tinsp.«,
or
Increase »FlowAcc«,
This can be set on-screen:
Set flow metering by means of »FlowAcc« on screen page
»Settings 1/1«.
and frequency f.
T
by
Assist
Paw
with plateau
T
plat
FlowAcc
P
plat
T
T
insp
e
1
T=
f
Flow
Operation
without plateau
FlowAcc
PEEP
T
insp
t
t
21