Pmax On/Off - Dräger Savina Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Savina:
Table of Contents

Advertisement

Display in information bar:
Replace flow monitoring
or
Replace O
monitoring
2
or
Replace flow monitoring and O
Flow monitoring is automatically activated each time Savina is
switched on.
When flow monitoring is deactivated, Savina does not
determine the following measured values:
— MV
— MVspont
Even when O
monitoring is deactivated, the devices requires
2
at least sensor 1 for accurate O
inaccurate O
/air mixes.
2
O
monitoring should only be deactivated when Savina is
2
supplied with air only.
O
monitoring remains deactivated even when the device
2
is switched on again.
Display in alarm line:
!! O
monitoring off
2
On acknowledging by pressing the »Alarm Reset« key,
display:
! O
monitoring off
2
Reactivate the monitor function after replacing the sensor.

Pmax ON/OFF

To use the pressure limit Pmax in the IPPV, IPPV
SIMV, SIMV/ASB ventilation modes.
The pressure limit value is set by means of the »Pinsp« key.
Savina limits the airway pressure to that set value.
∆∆
Press »Config.
« key repeatedly until
»Configuration 2/4« appears.
Select »Pmax« line = turn rotary knob,
to release for setting = press rotary knob.
To set »ON« = turn rotary knob,
confirm = press rotary knob.
The pressure limit »Pmax« is activated.
To set »OFF« = turn rotary knob,
confirm = press rotary knob.
The pressure limit »Pmax« is deactivated.
monitoring
2
control. Otherwise it delivers
2
and
Assist
Configuration
51

Advertisement

Table of Contents
loading