Technical data
Technical data
Environmental conditions
In operation
Temperature
Atmospheric pressure
Rel. humidity
In storage and transit
Temperature
Atmospheric pressure
Rel. humidity
Settings
Ventilation modes
Ventilation frequency f
Inspiration time T
insp
Tidal volume V
T
Accuracy
Inspiratory pressure P
insp
O
concentration
2
Accuracy
Positive end-expiratory pressure PEEP
or interm. PEEP
Trigger sensitivity
Pressure assist ∆ P
with PEEP
ASB
Flow Acceleration FlowAcc
*
BTPS
Body Temperature, Pressure, Saturated.
Measured values with reference to the conditions of the patient's lung,
o
body temperature 37
C, ambient pressure, steam-saturated gas.
** 1 mbar = 100 Pa
118
5 to 40°C
700 to 1060 hPa
5 to 95 %, no dew
–20 to 70°C
600 to 1200 hPa
10 to 95 %, no dew
IPPV/IPPVAssist,
SIMV, SIMV/ASB
BIPAP, BIPAP/ASB
CPAP/ASB
2 to 80 bpm
0.2 to 10 s
0.05 to 2.0 L, BTPS*
±10 % of set value or ±25 mL,
whichever is greater.
0 to 99 mbar**
21 to 100 vol.%
±3 vol.%
The accuracy of the inspiratory O
reduced when operating Savina without O
0 to 35 mbar
1 to 15 L/min
Can be switched off in IPPV ventilation mode.
0 to 35 mbar (relative to PEEP)
5 to 200 mbar/s
concentration is appreciably
2
sensors.
2