7.6.6 Submenu
You only obtain the following possible settings when the control variable is set to "External" (⇒ Section
7.7.4). See also 7.6.5.
Set control parameters
°C
O0c--
* Important:
The windows for setting the
additional parameters for the
external control for the Master
Controller:
Ds (=tve) and Dc (=tde),
and for the Slave Controller:
Da (=P-controlller),
are only displayed when
been changed from the works
setting of "Automatic" to
"Manual".
Changeover should only be
carried out by experienced
control technicians.
Continued on the
following page
YACE0072 / 21.08.07
(Master): Setting external control parameters
Prop.rng.master contr.[ ]
DO<4/
Res time master cntr [s]
Dl04/
Deriv time master cntr [s]
Ds013
Damping master cntr [s]
D
Dc0!3
has
P-controller [K]
Da 2/
Prop. range slave contr[K]
hO $/
°C
°C
°C
°C
°C
°C
Starting up
Display flashes. Set
with
or
.
Kpe range: 0,10 ... 9.80.
Display flashes. Set
with
or
.
Tne range: 5s ... 979s, Off.
*Display flashes. Set
with
or
.
Tve range: Off, 1s ... 800s.
*Display flashes. Set
with
or
.
Tde range: 0.0s ... 50.0s.
*Display flashes. Set
with
or
.
P-contr. range: 3K ... 250K.
Display flashes. Set
with
or
.
Xpf range: 0.3K ... 25K.
41