Lauda RP 845 Operating Instructions Manual page 42

Low-temperature thermostats with smartcool system
Table of Contents

Advertisement

Continuation of submenu
parameters
Return to the
previous page
* Important:
The windows for setting the
additional parameters for the
external control for the Master
Controller:
Ds (=tve) and Dc (=tde),
and for the Slave Controller:
Da (=P-controlller),
are only displayed when
been changed from the works
setting of "Automatic" to
"Manual".
Changeover should only be
carried out by experienced
control technicians.
42
LMDkt O0c--
Auto/man. for tne, tve, tde
D
D
has
Corr.variable limits [K]
Dr/K
Control works settings
cDE
End
Dlc
Starting up
Proline Low-temperature Thermostats
(Master): Setting external control
longer
than
@
°C
3 sec.
°C
longer
than
°C
3 sec.
°C
Display flashes. Changeover
tne, tve and tde from
@
LM
=Automatic to
=Manual
with
or
Display flashes. Set
with
or
.
Range: 0.0K ... 5.0K.
The works setting (default)
of the external control
parameters has been
cmlD
restored and
displayed.
YACE0072 / 21.08.07
.
is