LMDkt O0c--
7.6.5 Submenu
You only obtain the following possible settings when the control variable is set to "Internal" (⇒ Section
7.7.4. See also 7.6.6.)
Set control parameters
°C
O0c--
* Important:
The windows for setting the
internal control:
rs (=tv) and rc (= td)
are only shown when rsc
(=tvd) has been changed
from the works setting
"Automatic" to "Manual".
Changeover should only be
carried out by experienced
control technicians.
40
Proline Low-temperature Thermostats
(Master): Setting internal control parameters
Proportional range [K]
°C
^O &/
Reset time [s]
°C
rl 2/
Deriv. time [s]
°C
rs 10
Damping time [s]
°C
rc §4
Auto/manual for tv, td
°C
rsc @
Control works settings
°C
cDE
End
°C
Dlc
Starting up
Display flashes. Set
with
or
Xp range: 0.3K ... 25.0 K.
Display flashes. Set
with
or
Tn range: 3s ... 180s, Off.
*Display flashes. Set
with
or
Tv range: Off, 1s ... 145s.
*Display flashes. Set
with
or
Td range: 0.0s ... 25s.
Display flashes. Change tv
longer
@
and td from
=Autom. to
than
LM
=Manual
with
3 sec.
longer
The works setting (default)
than
of the internal control
parameters has been
cmlD
restored and
displayed.
3 sec.
.
.
.
.
or
.
is
YACE0072 / 21.08.07