Submenu Lmdkt B@K - Lauda RP 845 Operating Instructions Manual

Low-temperature thermostats with smartcool system
Table of Contents

Advertisement

Proline Low-temperature Thermostats
LMDkt B@K--
7.6.7 Submenu
temperature probe measuring chains (adjustment)
Calibrate temp. probe
°C
B@K--
YACE0072 / 21.08.07
(Master): Calibrating internal and external Pt100
Offset internal probe
°C
B@K 0
Int. Pt100 works settg.
°C
cDE 0
Offset external probe
°C
B@K D
Ext. Pt100 works settg.
°C
cDE D
End
°C
Dlc
Starting up
longer
than
Display flashes. Use
or
to enter value
of reference thermometer.
3 sec.
longer
The works setting (default)
than
for the offset for the internal
probe has been restored
cmlD
and
is displayed.
3 sec.
longer
than
Display flashes. Use
or
to enter value
of reference thermometer.
3 sec.
longer
The works setting (default)
than
for the offset for the external
probe has been restored
cmlD
and
is displayed.
3 sec.