Lauda RP 845 Operating Instructions Manual page 23

Low-temperature thermostats with smartcool system
Table of Contents

Advertisement

Proline Low-temperature Thermostats
Hose connections
a) Elastomer hoses
Hose type
EPDM hose
uninsulated
EPDM hose
uninsulated
EPDM hose
insulated
Silicone hose
uninsulated
Silicone hose
insulated
Viton
Viton
insulated
Viton
insulated
b) Metal hoses in non-rusting stainless steel with union nut M 16x1, internal width 10 mm.
Type
MC 50
MC 100
MC 150
MC 200
MK 50
MK 100
MK 150
MK 200
YACE0072 / 21.08.07
Internal width
Ø mm
9
12
12
External Ø.
approx. 35mm
11
11
External Ø.
approx. 35mm
11
8.5
External Ø.
approx. 30mm
11
External Ø.
approx. 32mm
− EPDM hose is not suitable for Ultra 350 and not suitable for mineral oils.
− With silicone rubber, silicone oils lead to substantial swelling
silicone oil with silicone hoses.
− Secure hoses against slippage with hose clips.
Length
Temperature range °C
(cm)
50
10...400
100
10...400
150
10...400
200
10...400
50
-90...200
100
-90...200
150
-90...200
200
-90...200
Temperature
range °C
10...120
10...120
-60...120
-30...100
-60...100
10...200
-60...150
-60...150
Field of application
With single insulation,
for all heat carrier liquids
With foam insulation for
refrigeration range,
for all heat carrier liquids
Preparation
Field of
Order number
application
For all heat carrier
liquids except
Ultra 350 and
mineral oils
For all heat carrier
liquids except
Ultra 350 and
mineral oils
For all heat carrier
liquids except
Ultra 350 and
mineral oils
Water
Water/ glycol
mixture
Water
Water/ glycol
mixture
For all heat carrier
liquids
For all heat carrier
liquids
For all heat carrier
liquids
Order number
"
"
"
"
"
"
RKJ 111
RKJ 112
LZS 021
RKJ 059
LZS 007
RKJ 091
LZS 017
LZS 018
never use
LZM 040
LZM 041
LZM 042
LZM 043
LZM 052
LZM 053
LZM 054
LZM 055