[Sense[1]]|Sense2:Average[:State] <Boolean - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage[:STATe] <Boolean>
[SENSe[1]]|SENSe2:AVERage[:STATe] <Boolean>
This command is used to enable and disable averaging.
Syntax
Example
Reset Condition
On reset, averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of
averaging.
Query Example
SENS2:AVER?
Error Message
If [SENSe[1]]|SENSe2:AVERage:STATe is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, "Settings
Conflict" occurs.
9-12
:
SENS
1
SENS2
AVER 1
1 is returned when averaging is enabled.
0 is returned when averaging is disabled.
:STAT
AVER
This command enables averaging
on channel A.
This command queries whether
averaging is on or off for channel B.
HP E4418B/E4419B Programming Guide
Space
0|OFF
1|ON
?