MEMory Subsystem
MEMory:CATalog:TABLe?
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these sensor calibration tables is based on statistical
averages for a range of Hewlett-Packard power sensors. These tables can
be altered. The predefined data is listed in your User's Guide. These power
sensors are:
1. Default is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 and ten frequency dependent offset tables named
CUSTOM_A through CUSTOM_J which do not contain any data when the
power meter is shipped from the factory.
Syntax
Example
7-8
Table
0
1
2
HP 8482A, HP 8482B, HP 8482H
3
4
5
6
7
8
9
:CAT
MEM
MEM:CAT:TABL?
Power Sensor
None
HP 8481A
HP 8483A
HP 8481D
HP 8485A
HP R8486A
HP Q8486A
HP R8486D
HP 8487A
:TABL
?
This command queries the list of
stored tables.
HP E4418B/E4419B Programming Guide
Table Name
1
DEFAULT
HP8481A
HP8482A
HP8483A
HP8481D
HP8485A
R8486A
Q8486A
R8486D
HP8487A