Status:questionable:calibration[:Summary] - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

STATus:QUEStionable:CALibration[:SUMMary]

The questionable calibration summary register set contains bits which
give an indication of the quality of the data produced by the power meter
due to its calibration status.
The following bits in these registers are used by the power meter:
These bits are set by the following:
These bits are cleared when any of the three commands listed above
succeed and no errors are placed on the error queue.
Syntax
HP E4418B/E4419B Programming Guide
Bit
Decimal
Definition
Number
Weight
0
-
Not used
1
2
Summary of Channel A CALibration
2
4
Summary of Channel B CALibration (HP E4419B only)
3-15
-
Not used
Error -231, "Data questionable; CH<A|B>:ZERO ERROR"
Error -231, "Data questionable; CAL ERROR"
Error -231, "Data questionable; CAL ERROR ChA"
Error -231, "Data questionable; CAL ERROR ChB"
:QUES
STAT
STATus:QUEStionable:CALibration[:SUMMary]
:CAL
:SUMM
STATus Subsystem
10-21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4419b