Questionable Status Register
The questionable status register provides information about the quality of
the power meter's measurement results. Any or all of these conditions can
be reported in the questionable data summary bit through the enable
register. You must write a value using the
STATus:QUEStionable:ENABle command to set the enable register
mask.
The questionable status model is shown in the pullout at the end of this
chapter.
The following bits in these registers are used by the power meter.
Bit
Decimal
Number
Weight
0 to 2
-
3
8
4 to 7
-
8
256
9
512
10 to 14
-
15
-
HP E4418B/E4419B Programming Guide
Definition
Not used
POWer Summary
Not used
CALibration Summary
Power On Self Test
Not Used
Not used (always 0)
Power Meter Remote Operation
Status Reporting