Read[1|2][:Scalar][:Power:ac]? [<Expected_Value>[,<Resolution>[,<Source List>]]] - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

MEASurement Instructions
READ[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
READ[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to single
channel with relative mode off, aborts then initiates the specified channel,
calculates the measurement result and places the result on the bus. The
result is a power based measurement and is expressed in the units defined
by UNIT[1|2]:POWer.
Note
Syntax
READ
1
:SCAL
2
Parameters
Refer to "The Optional Parameters", on page 2-3 for additional details on
the parameters in this command.
2-36
INITiate:CONTinuous must be set to OFF, otherwise error -213,
"INIT ignored" occurs. If TRIGger:SOURce is set to BUS, error
-214, "Trigger deadlock" occurs.
:POW
expected_value
Space
DEF
Item
Description/Default
expected_value
The expected power level
(for the
parameter can be set to DEF or a
expected
numeric value. If a value is
power level)
entered it should correspond to
that set by CONFigure otherwise
an error occurs.
:AC
?
,
resolution
DEF
HP E4418B/E4419B Programming Guide
,
source list
Range of Values
sensor dependent
1
DEF

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4419b

Table of Contents