CONFigure[1|2][:SCALar][:POWer:AC]:DIFFerence [<expected_value>[,<resolution>[,<source list>]]]
CONFigure[1|2][:SCALar][:POWer:AC]:DIFFerence
[<expected_value>[,<resolution>[,<source list>]]]
This command is only relevant for the HP E4419B power meter.
This command is used to set the measurement function and resolution of
the specified window. It sets the measurement function to difference with
relative mode off.
Syntax
CONF
1
2
Parameters
Refer to "The Optional Parameters", on page 2-3 for additional details on
the parameters in this command.
HP E4418B/E4419B Programming Guide
:POW
:SCAL
expected_value
Space
DEF
Item
Description/Default
expected_value
The power meter ignores the
(for the
numeric value entered in this
expected
parameter. Any value entered is
power level)
treated like DEF.
resolution
A numeric value for the resolution.
If unspecified the current
resolution setting is used.
:DIFF
:AC
,
resolution
DEF
MEASurement Instructions
,
source list
Range of Values
sensor dependent
1
DEF
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF
2-13