MEASurement Instructions
MEASure[1|2][:SCALar][:POWer:AC]:RATio:RELative? [<expected_value>[,<resolution>[,<source
list>]]]
MEASure[1|2][:SCALar][:POWer:AC]:RATio:RELative?
[<expected_value>[,<resolution>[,<source list>]]]
This command is only relevant for the HP E4419B power meter.
This command sets the specified window's measurement function to ratio
mode with relative mode on, aborts, configures then initiates both channel
A and B, calculates the ratio measurement and places the result on the
bus. The result is a ratio based measurement and is expressed in the units
defined by UNIT[1|2]:POWer:RATio. The relative value used is that set
by the CALCulate:RELative:MAGNitude:AUTO command.
Syntax
MEAS
1
2
Parameters
Refer to "The Optional Parameters", on page 2-3 for additional details on
the parameters in this command.
2-60
:POW
:SCAL
expected_value
Space
DEF
Item
Description/Default
expected_value
The power meter ignores the
(for the
numeric value entered in this
expected
parameter. Any value entered is
power level)
treated like DEF.
resolution
A numeric value for the resolution.
If unspecified the current
resolution setting is used.
:RAT
:REL
:AC
,
resolution
DEF
HP E4418B/E4419B Programming Guide
?
,
source list
Range of Values
sensor dependent
1
DEF
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF