MEMory Subsystem
MEMory:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>}
MEMory:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>}
This command is used to enter calibration factors into the sensor
calibration table, or offsets into the frequency dependent offset table,
currently selected using MEMory:TABLe:SELect. Any previous
calibration factor list, or offset list is cleared before the new calibration
factors/offsets are stored.
A maximum of 81 parameters for sensor calibration tables and 80
parameters for frequency dependent offset tables can be sent with this
command. For sensor calibration tables only, the first parameter is the
reference calibration factor, each subsequent parameter is a calibration
factor point in the sensor calibration table.
Entries in the frequency lists correspond as shown with entries in the
calibration/offset factor lists.
For sensor calibration tables the number of frequency points must be one
less than the number of calibration factor data points. This is verified
when the sensor calibration table is selected using
SENSe:CORRection:CSET1:SELect <string>.
Syntax
MEM
7-26
Calibration Factor/Offset
Reference Calibration Factor
(Sensor Cal Table Only)
Calibration Factor/Offset 1
"
Calibration Factor/Offset 80
:TABL
:GAIN
Frequency
Frequency 1
Frequency 80
Space
:MAGN
HP E4418B/E4419B Programming Guide
-
"
,
numeric_value
?