SYSTem:PRESet
This command is used to preset the power meter. The result of this
command is similar to the *RST command. Refer to Table 11-1 for more
information about the preset configuration.
Note
Syntax
Example
Command
CALC:GAIN[:MAGN]
CALC:GAIN:STAT
CALC:LIM:CLE:AUTO
CALC:LIM:FAIL
CALC:LIM:FCO
CALC:LIM:LOW[:DATA]
CALC:LIM:STAT
CALC:LIM:UPP[:DATA]
CALC:MATH[:EXPR]
CALC:REL[:MAGN]:AUTO
CALC:REL:STAT
CAL:AUTO
HP E4418B/E4419B Programming Guide
All settings apply to both *RST and to SYSTem:PREset unless
stated otherwise.
:PRES
SYST
SYST:PRES
Table 11-1: Preset Settings
Setting
0.000 dB
OFF
ON
OFF
OFF
-90 dBM or -90 dB
OFF
+90 dBM or +90 dB
HP E4418B:
Upper - channel A
Lower - channel A
HP E4419B
Upper - channel A
Lower - channel B
OFF
OFF
OFF
SYSTem Subsystem
This command presets the
power meter.
Comments
Display offset value
Display offset disabled
Clear limit data at INIT
No failures
Limit failure count is zero
Lower limit
Window limits checking disabled
Math expression
Reference value disabled
Relative offset disabled
Automatic calibration disabled
SYSTem:PRESet
11-29