[Sense[1]]|Sense2:Power:ac:range:auto <Boolean - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <Boolean>
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <Boolean>
This command is used with the HP E-series power sensors to enable and
disable autoranging. When autoranging is ON, the power meter selects the
best measuring range for the measurement. When autoranging is set to
OFF, the power meter remains in the currently set range.
The [SENSe[1]]|SENSe2:POWer:AC:RANGe command disables
autoranging.
If INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to
IMMediate, the range will track the input power if
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO is ON.
If the power meter is not making measurements then autoranging only
occurs when the power meter is triggered.
Syntax
SENS
1
SENS2
Example
Reset Condition
On reset, autoranging is enabled.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO?
The query enters a 1 or 0 into the output buffer indicating the status of
autoranging.
9-52
:
POW
:AC
POW:AC:RANG:AUTO 0
1 is returned when autoranging is enabled.
0 is returned when autoranging is disabled.
:AUTO
:RANG
This command disables
autoranging.
HP E4418B/E4419B Programming Guide
Space
0|OFF
1|ON
?

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4419b

Table of Contents