Example
In case of a device error, the red LED will repeat a flash sequence that corresponds to the particular type
of error that occurred. For example, in case of a RAM checksum error (code=4), the LED will show a
sequence of 4 flashes followed by a 750 millisecond pause. After the pause the flashing starts over again
in the same manner. This flash/pause sequence will be continuously repeated.
RAM checksum error (code 4) appears with the following Error LED (red) flash sequence as illustrated in
Figure
Figure 5-3. Example of an error code flash sequence