4. Choose the Common Settings option. This option lets you apply the same level rate threshold for all
servo tanks ("servo tank" means that Tank Type is set to one of the servo tank options available in the
tank installation wizard).
5. Enter the desired Hysteresis and Threshold values. The Hysteresis variable lets you specify a zone in
which Fast Polling is not changed from its current state. This prevents switching Fast Polling on and off
when the level rate is changed only small amounts around a certain Threshold value.
6. Click Apply to save the current configuration. Click OK to close the window.
Fast Polling
Fast Polling
March 2017
187