Emerson Rosemount 2460 Reference Manual page 190

System hub for tank gauging systems
Hide thumbs Also See for Rosemount 2460:
Table of Contents

Advertisement

®
Modbus
Standard Question
March 2017
Register
Register Name
Number
4224-4351
User Defined Value Status
4352-4479
Temperature Element Status
4608-4671
Temperature Element in Product
Status
4672-4735
Temperature Element in Vapor
Status
4736-4799
Temperature Element Blocked
Status
180
Type
Unit
DWORD
Bitfield
DWORD
Bitfield
WORD
Bitfield
WORD
Bitfield
WORD
Bitfield
Reference Manual
00809-0100-2460, Rev DB
Description
Priority 3 question
Bit 0: Simulated User Def 1
Bit 1: Manual User Def 1
Bit 2: Invalid User Def 1
Bit 3: CFail User Def 1
Bit 4: Backup User Def 1 used
Bit 5: Spare
Bit 6: Simulated User Def 2
Bit 7: Manual User Def 2
Bit 8: Invalid User Def 2
Bit 9: CFail User Def 2
Bit 10: Backup User Def 2 used
Bit 11: Spare
Bit 12: Simulated User Def 3
Bit 13: Manual User Def 3
Bit 14: Invalid User Def 3
Bit 15: CFail User Def 3
Bit 16: Backup User Def 3 used
Bit 17: Spare
Bit 18: Simulated User Def 4
Bit 19: Manual User Def 4
Bit 20: Invalid User Def 4
Bit 21: CFail User Def 4
Bit 22: Backup User Def 4 used
Bit 23: Spare
Bit 24: Simulated User Def 5
Bit 25: Manual User Def 5
Bit 26: Invalid User Def 5
Bit 27: CFail User Def 5
Bit 28: Backup User Def 5 used
Bit 29: Spare
Priority 3 question
Bit 0: Invalid value E1
..
Bit 15: Invalid value E16
Bit 16: Cfail E1
..
Bit 31: Cfail E16
Bit 0: Element1 in product
..
Bit 15: Element16 in product
Bit 0: E1 in vapor
..
Bit 15: E16 in vapor
Blocked from Avg. calculation
Bit 0: E1 is blocked.
..
Bit 15: E16 is blocked
®
Modbus
Standard Question

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents