Weinmann WM 13011 Description And Operating Instructions page 44

Oxymat 3 oxygen concentrator from serial number 8000
Table of Contents

Advertisement

108
6. To check flowmeter
the connecting tube
ing tube
off connection
17
ter and hold connection
closed. The flowmeter is free from leaks if
the flowmeter ball
the bottom.
7. Check mains failure alarm. To do this,
pull out the mains plug while the
OXYMAT 3 is switched on. An acoustic
signal is heard, and the red fault
indicator
lights up.
18
8. Insert the mains plug in the power socket
again. The alarm signal should stop, and
the device starts again.
9. Switch off your OXYMAT 3 when you no
longer need it.
English
for leaks, bend
16
. Or pull connect-
17
on flowme-
28
28
sinks all the way to
22

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 13021Wm 13031Wm 13041Wm 13017Wm 13018