Changing Patients - Weinmann WM 24042 Description And Operating Instructions

Oxygen valve
Table of Contents

Advertisement

Available languages

Available languages

Disinfect creased tube and adapter
Fit double tube
Branched end
of double tube
Grey adapter
Double tube
Sleeve
Transparent adapter
After disinfecting
Dry the assembled tube system completely with the aid of the therapy device. The proce-
dure can be found in the instructions for use for your therapy device.

5.3 Changing patients

To prepare the device for use by a new patient, replace the O
system and clean the outside surfaces of the O
1. Proceed as described in "Clean the tube system".
2. After disinfecting, rinse thoroughly with distilled water.
1. Push the short part of the double tube through the
transparent adapter and then push it as far as possible
onto the gray adapter.
2. Push the sleeve back onto the gray adapter.
3. Push the long part of the double tube into the groove
of the transparent adapter. The two ends of the tube
must lie smoothly next to one another.
4. Insert the free end of the double tube in the creased
tube.
5. Push the sleeve of the creased tube onto the
Sleeve
transparent adapter.
supply hose or the hose
2
valve by wiping them with disinfectant.
2
Hygienic preparation
41
EN