Operation; Before Switching On; Switching On And Off - Weinmann WM 13011 Description And Operating Instructions

Oxymat 3 oxygen concentrator from serial number 8000
Table of Contents

Advertisement

O / I
90
O/ I
O
2

5. Operation

5.1 Before switching on

If the device has been stored in a cold en-
vironment (below +10 °C) before being
switched on, put it in a warm room and
wait approximately 30 minutes before
switching it on to allow the OXYMAT 3 to
come to room temperature.
Unscrew humidifier glass
humidifier from dosage monitor
glass with distilled (aqua dest. from your
pharmacy), demineralised or boiled wa-
ter up to the mark. Then screw humidifier
glass on again.
If you use a sterile water system, fit this in
accordance with the appropriate installa-
tion instructions.

5.2 Switching on and off

1. Switch on the OXYMAT 3 using On/Off
knob
. A self-test will then run.
2
– Red fault indicator
dicator
and green power indicator
19
immediately come on briefly.
20
– Green power indicator
es for the duration of the self-test (ap-
prox. 1 minute).
– During the self-test you will hear the
compressor stopping and system pres-
sure being released after approx.
30 seconds (hissing noise). This is an
internal leak test taking place. The
compressor then starts up again.
– If the device is functioning perfectly,
only green power indicator
be on after the self-test.
English
of bubble
25
and fill
9
, yellow status in-
18
then flash-
20
will still
20