    2. Description of the resuscitator ....7 2.1 COMBIBAG components ....7 3.
  • Page 4: Basic Principle

    1. Basic principle 1.1 Purpose The COMBIBAG is for the manual ventilation of chil- dren (from a body weight of 10 kg) and adults via a face mask or endotracheal tube. The respiratory air can be enriched with oxygen. 1.2 Function...
  • Page 5: Special Features Of The Combibag

    Now the special design featuring the COMBIBAG covers both functions. The bag is divided into two segments (one large and one small) by two longitu- dinal troughs, marked with appropriate symbols.
  • Page 6 With the COMBIBAG 2-stage safety valve it is now possible to set a pressure limit: •...
  • Page 7: Description Of The Resuscitator

    > 20cm H pressure < 20cm H effect 2. Description of the resuscitator 2.1 COMBIBAG components The COMBIBAG resuscitator for adults and children consists of 4 working parts: Patient valve Patient valve (non- rebreathing walve) The patient valve consists of: Safety valve •...
  • Page 8 • intake port cover • valve plate to allow the inlet of fresh gas and forming a seal when the bag is squeezed • silicone ring for hanging up the COMBIBAG. Description of the resuscitator...
  • Page 9: Safety Instructions

    • The detailed knowledge and attention of this operating instructions are the basis for proper use of the COMBIBAG. The instrument is just de- termined for the field of use stated. • Malfunctions and a lack of biocompatibility may result if third-party articles are used.
  • Page 10 • More copies of this operating instructions are available from the manufacturer WEINMANN Emergency. Safety instructions...
  • Page 11: Operating Instructions

    “rattling” are covered. 4.2 Checking the function The function of the COMBIBAG resuscitation bag should be proven before each use and can be checked in a few seconds as follows: •...
  • Page 12: Guide Values For Ventilation

    • Relief valve setting 20 mbar (2 kPa (≈ 20 cmH O)): on closing the inhalation pas- sage with the ball of the left thumb and squeez- ing the bag with the right hand, air flows through the relief valve openings. •...
  • Page 13: Mask Ventilation

    • lay the patient on his back • extend the head and neck • pull the red ring on the COMBIBAG relief valve upwards (pressure limit 20 mbar (2 kPa (≈ 20 cmH O))) • C-grip: left hand lifts the lower jaw back and seals the mask over the mouth and nose •...
  • Page 14: Ventilation Via An Endotracheal Tube

    • Intubation • Press the red ring on the COMBIBAG relief valve downwards (pressure limit 60 mbar (6 kPa (≈ 60 cmH O))) • Even with ventilation via an endotrachel tube, one should aim for a pressure limit of 20 mbar (2 kPa (≈...
  • Page 15: Peep Ventilation

    4.8 Volume measurement To prevent hypo- or hyperventilation in the case of prolonged manual resuscitation with the COMBIBAG, a volumeter for measuring the expired gas volume can be connected to the expiration branch of the patient valve. Operating instructions...
  • Page 16: Use Of Hme-Filters

    4.9 Use of HME-Filters For hygienical reasons and for climatizing the breathing air, the inspiratory branch of the patient valve can be fitted with commercial 15/22 mm stan- dard cone HME-filters (heat and moisture exchang- er). This insignificantly increases the inspiratory as well as the expiratory resistance.
  • Page 17: Hygienic Preparation

    Upon completion of the purification process, carry out a functional test. Patient valve Individual parts of the COMBIBAG resuscitator are made of silicon or polysufone. Safety valve clean Use soapy water to COMBIBAG and rinse un- (relief valve) der running water.
  • Page 18: Cleaning During Use

    Checking the function” on page 11). 5.2 Maintenance and service With proper use and maintenance of the COMBIBAG no additional service is required. Just the intake diaphragm should be examined once a year. The diaphragm is ready for proper use if no cracking spots are visible.
  • Page 19: Spare Parts

    6. Spare parts 6.1 Standard scope of supply Item no.* Description of component Order no. COMBIBAG resuscitation bag for adults and WM 11090 children, without face masks COMBIBAG resuscitation bag for adults and WM 11020 children with 1 face mask, consisting of: –...
  • Page 20: Accessories

    Item no.* Description of component Order no. COMBIBAG resuscitation bag for adults and WM 11050 children with 3 face masks, consisting of: – COMBIBAG resuscitation bag WM 11000 – ventilation mask with inflatable silicone WM 5074 cushion for adults, Size 5 –...
  • Page 21: Spare Parts

    *See illustration on page 23. 6.3 Spare parts For COMBIBAG Item no.* Description of component Order no. COMBIBAG resuscitation bag for adults and WM 11000 children Cord ring, Ø 42 mm (old suspension ring) WM 1145/43 Suspension ring, new, Ø 60 mm WM 1145/49 Seal 30/35 x 0.5 mm...
  • Page 22 Item no.* Description of component Order no. Bellows WM 11001 Valve housing WM 11002 6 – 12 Ventilation valve, without safety valve WM 11003 Guide stub WM 11004 Valve element WM 11006 Valve insert WM 11007 Inlet membrane WM 11008 Intake stub, inner part WM 11021 Intake stub, outer part...
  • Page 23 Abbildung Spare parts...
  • Page 24: Technical And Performance Data

    7. Technical and performance data COMBIBAG Classification acc. to EC directive 93/42/EEC Dimensions (L x D) appr. 330 mm x 125 mm (inflated) Weight appr. 410 g Scope of use: – Child grip 10 – 16 kg body weight – adult grip >16 kg body weight...
  • Page 25 Tidal volumes and respiratory frequencies (cycles per minute) The follwing frequencies and tidal volumes are ob- tained during ventilation with COMBIBAG: Cycles per min. Tidal volume in ml Child grip 2 fingers 100 - 150 3 fingers > 60 150 - 200...
  • Page 26: Warranty

    In the event of a claim under warranty, contact your specialist dealer. Product Warranty period WEINMANN Emergency devices including accessories (except masks) for 2 years oxygen medicine and emergency medicine Masks including accessories, rechargeable batteries, batteries (unless quoted 6 months...
  • Page 28 Manufacturer Center for Production, Logistics, Service WEINMANN Emergency WEINMANN Emergency Medical Technology GmbH + Co. KG Medical Technology GmbH + Co. KG Frohbösestraße 12 Siebenstücken 14 22525 Hamburg 24558 Henstedt-Ulzburg GERMANY GERMANY T: +49 40 88 18 96-120 E: customerservice@weinmann-emt.de...