3.11 AC Current Probe MR193 Probe

When installing probes, face the arrow on the probe in the direction of
the load.
Nominal Range: 1000A
Measurement Range: 10A to 1000A
Currents <1A
Probe Output Signal: 1mV/A
Maximum Cable Diameter: One 1.6" (42mm) or two 0.98" (25.4mm) or
two bus bars 1.96 x 0.19" (50 x 5mm)
Safety: EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III
Reference Conditions:
Ambient temperature
Humidity
Battery voltage
Position of conductor in the sensor
Magnetic field
AC External magnetic field
External electric field
Frequency
Type of signal measured
26
)
AC
(% of the output signal)
-4 to 140°F (-20 to 60°C)
10 to 90% RH
10Hz to 20kHz
Conductor in contact with
, 1400A
AC
will be displayed as zero with this probe.
AC/DC
20 to 100A
≤3%
Range
Any position
the sensor
(3945-B accuracy included)
max
DC
10A to 1300A
,
AC
64 to 82°F (18 to 28°C)
Power Quality Analyzer Model 3945-B
100 to 1000A
-90° ±0.5°
Error
0.2% per 10°C
0.5%
0.5%
< 2%
(6% near latching system)
<0.7%
(6% near latching system)
PEAK AC+DC
20 to 75% of RH
9V ±0.1V
Centered
DC magnetic field
None
None
≤65Hz
Sinusoidal
≤2%